How to Use a Lab Autoclave: Loading, Cycles and Safe Unloading

November 21, 2026

Last updated November 2026 · LORDERAN how-to guide

Short answer: fill the chamber with distilled water to just below the shelf, load wrapped items with gaps for steam and liquids no more than three-quarters full under cotton plugs, and pick the right program: no exhaust and no drying for liquids, exhaust and drying for solids. A common cycle is 121 °C (250 °F) for 30 minutes. Open the lid only when the pressure gauge reads zero, put an indicator in every load and run a spore test regularly. The figures come from the LORDERAN vertical autoclave manual (35 to 100 L) and the CDC sterilization guideline.

1. Install and fill

The vertical autoclave makes its own steam from water in the bottom of the chamber, so the water level and water quality matter every cycle.

Where it goes

  1. 1
    Level, firm floorWith 60 cm (24 in) clear from the walls.
  2. 2
    The right room5 to 40 °C (41 to 104 °F) and no more than 85 %RH.
  3. 3
    Its own grounded outletA dedicated outlet with a clearly marked disconnect, at the voltage on the rating plate (220 V, or 110 V 60 Hz versions). Never remove the ground wire.
  4. 4
    Drain hose end in the airThe outlet of the drain hose must never sit under water, or water can be drawn back into the chamber.

Water

  1. 1
    Distilled waterTap water is not recommended.
  2. 2
    About 5 L, below the shelfThe level reaches the foot of the chamber, under the shelf. Top up after every cycle when you run several in a row.
  3. 3
    Auto-fill modelsPut the inlet hose to the bottom of a 5-gallon bucket of water. Filling stops when the water light turns steady green.
  4. 4
    DrainingWait at least 2 hours after a cycle, with the lid open, before you drain hot water.

2. Load it

  1. 1
    Leave room for steamWrap items properly and leave gaps between them. Center the basket; stack baskets one above the other.
  2. 2
    LiquidsHeat-resistant glass bottles, no more than ¾ full, closed with a cotton plug. No solid rubber stoppers or corks.
  3. 3
    Open vesselsBeakers, flasks and tubes go upside down or on their side, so air can drain out and steam can get in.
  4. 4
    One kind per loadDo not mix dressings and liquids in the same load.
  5. 5
    An indicator in every loadPut a chemical indicator card in the load.
Never load flammable, explosive or oxidizing materials, or media that corrode stainless steel. Keep the safety valve vent clear.
LidSafety valve: keep clearWater, about 5 L, below the shelfBasketLiquids ≤ ¾ full,cotton plugOpen vesselsupside downWrapped items,gaps betweenGrey dashes: steam rising through the load

3. Choose the cycle

The controller sets 116 to 134 °C and 4 to 120 minutes of sterilizing time, with up to 240 minutes of drying. CDC gives 121 °C (250 °F) for 30 minutes as the minimum exposure for wrapped items in a gravity-displacement autoclave like this one.

  1. A
    Program A: liquidsSterilizes only, with no automatic venting and no drying. Use it for media and solutions.
  2. C
    Mode C: solidsSterilizes, exhausts the steam and dries. Use it for wrapped instruments, glassware and dry goods. Never use drying for liquids.
  3. ✓
    Air comes out firstThe autoclave vents the cold air in three pulses between about 105 and 115 °C before it climbs to the set temperature, because trapped air keeps steam away from the load.
  4. !
    Leave the venting settings aloneThe end-of-exhaust temperature can be set from 110 to 120 °C. Change it only when you need to.
Heat and purge airSterilize: 121 °CExhaust and dry20601001213 air-purge pulsese.g. 30 min°C

Shape of a typical Mode C cycle. Heat-up and drying times depend on the load.

4. Run and unload

The end of the cycle, when the lid opens, is where pressure and heat can hurt you.

Close and start

  1. 1
    Close the lidSlide the crossbeam to the left column and turn the handwheel clockwise until the door light comes on, then on until sealed.
  2. 2
    StartCheck the program and the time, then start. Do not touch the lid during the cycle.

Wait for zero

  1. 1
    Gauge at zeroOpen the lid only after the pressure gauge reads zero. Forcing it earlier damages the lock.
  2. 2
    LiquidsNever vent a liquid load. Wait for zero, then let the liquid cool before you take the bottles out.
  3. 3
    After dryingIf the lid holds by vacuum, lift the relief valve bonnet to let air in.

Unload

  1. 1
    Gloves onWear heat-protective gloves for hot items.
  2. 2
    Read the indicatorCheck the indicator card before you use the load.
  3. 3
    Take it out promptlyThe autoclave is not for storage. Leave the lid open 5 minutes before the next cycle.
If pressure goes above 0.24 MPa and the safety valve does not open: cut the power, open the manual vent, wait until the gauge has read zero for 1 to 2 minutes, then open the lid. The safety valve must be replaced before the next use.

5. Prove it worked

The temperature reading shows the chamber got hot; indicators show the conditions reached the load.

  1. 1
    Every loadA chemical indicator, checked before the load is used.
  2. 2
    At least weeklyA biological indicator (spore test), as CDC recommends for steam sterilizers.
  3. 3
    After installation, a move or a major repairThree consecutive cycles with biological and chemical indicators before routine use, and the same after a sterilization failure.
  4. 4
    A positive spore testRepeat the test straight away. If it stays positive, stop using the autoclave until it has been serviced.

Error codes

Cut the main power before you troubleshoot. Most repairs need qualified service personnel. The alarm light also flashes when the waste water tank is full.

CODEMEANINGWHAT TO DO
ERR-1The lid is not closed.Close the crossbeam and turn the handwheel until sealed.
ERR-2Water below the low level.Add water to the foot of the chamber.
ERR-3Sterilizing temperature too high.Stop and call service.
ERR-5Heating element overheat protection.Switch off, let it cool, check the water level; call service if it repeats.
ERR-6Pressure too high.Stop the cycle; call service. See the warning above.
ERR-7 / ERR-8Power cut, or the cycle was stopped by hand.Unlock with the function key and decide whether to rerun the load.
ERR-9 / ERR-10Chamber or wall temperature sensor fault.Call service.
ERR-12Drying heater fault.Call service.
ERR-20Heating took too long.Check the water level and the supply voltage; call service if it repeats.
ERR-21Sterilization failed.Treat the load as not sterile and run it again.
ERR-30Auto water fill failed (auto-fill models).Check the water bucket, inlet hose, pump and inlet valve.

Maintenance

Disconnect the power and let the autoclave cool before any cleaning. Before moving it, drain the chamber, close the lid and switch off.

EVERY USEWater and ventCheck the water level and that the safety valve vent is clear.
WEEKLYExercise the safety valveLift and lower its handle several times so scale cannot seize it.
MONTHLYDescale and cleanDrain, cool, take out shelf and baskets, brush scale off walls, heater and probes, wipe with alcohol or distilled water.
QUARTERLYDoor lockCheck that the self-locking mechanism moves freely.
EVERY 6 MONTHSGauge and greaseCompare the pressure gauge with a reference gauge; grease the crossbeam screw with high-temperature grease.
YEARLYGasket and inspectionReplace the lid gasket. Have the safety valve and the vessel inspected by an authorized body.

Manuals and guides

Each LORDERAN autoclave has its own manual with its programs, limits and error list.

35 to 100 L, with dryingVertical autoclave manual
Every modelAutoclave documents
Buying guideHow to choose an autoclave
All modelsCollection

Questions

Is 121 °C for 30 minutes always right?

It is the CDC minimum for wrapped items in a gravity-displacement autoclave. Large liquid volumes and dense loads take longer to reach temperature, so validate your own loads with indicators.

Can I dry a liquid load?

No. Use Program A for liquids: it does not vent or dry. Drying is for solids only.

Why won't the lid open?

The chamber is still under pressure, or, after drying, under vacuum. Wait for the gauge to read zero; after drying, lift the relief valve bonnet to let air in.

Can I use tap water?

The manual specifies distilled water and does not recommend tap water. Clean out scale from the walls, heater and probes every month either way.

How full can I fill bottles?

No more than three-quarters, in heat-resistant glass with a cotton plug. Solid rubber stoppers and corks are not allowed.

How often should I run a spore test?

CDC recommends a biological indicator at least weekly, plus three consecutive cycles with indicators after installation, a move, a major repair or a failure.