How to Use a Lab Centrifuge: Setup, Balancing and a Safe Run

November 21, 2026

Last updated November 2026 · LORDERAN how-to guide

Short answer: set the centrifuge on a solid, level bench with 300 mm (about 12 in) clear on every side. Seat the rotor on a clean taper and tighten the nut. Load tubes symmetrically, with opposite positions within 0.1 g of each other. Fill to no more than 90 % in swing-out rotors and 80 % in fixed-angle rotors. Then select the rotor, set × g, time and temperature, close the lid until it clicks and stay until the run is at speed and quiet. The figures in this guide come from the LORDERAN TDL-5A and TGL-21M manuals. Check the manual for your model; all ten are linked at the end.

1. Place it

A centrifuge pushes warm air out of its vents and needs room to shed it. The clear zone is also a safety zone: nobody should lean on the instrument or stand beside it longer than needed while it runs.

  1. 1
    Solid, level benchIt must carry the instrument, rotor and load without flexing, with every rubber foot touching: at least 80 kg (176 lb) plus load for the TDL-5A, 100 kg (220 lb) for the TGL-21M.
  2. 2
    300 mm (12 in) clear on every sideAway from pumps, shakers and other vibration, direct sun and heat sources.
  3. 3
    Its own grounded outletAt the voltage on the rating plate. No extension cords or power strips.
  4. 4
    Two hours to acclimateAfter cold transport or storage, wait at least 2 hours before switching on so condensation can dry.
SOLID, LEVEL BENCH300 mm300 mm · front300 mm300 mmWarm air from the ventsKeep clear: cooling air and safety zone
Drive housingTaper: wipe clean,thin film of lubricantLower it level and straight downNut: clockwise

2. Fit the rotor

A rotor that is not fully seated, or a loose nut, is a common cause of vibration. The fit takes a minute and is worth checking before every run.

  1. 1
    Clean the taper and the boreWipe both with a soft cloth and apply a very thin film of the supplied lubricant.
  2. 2
    Lower it straight downHold the rotor level and let the bore seat fully on the taper. Never drop or force it.
  3. 3
    Tighten the nut clockwiseHold the rotor still with the other hand. On fixed-angle rotors, fit and tighten the rotor lid too.
  4. 4
    Hang every bucketOn swing-out rotors all buckets go on, empty ones included, and each must swing freely.

To remove it, switch off first, hold the rotor, loosen the nut counterclockwise and lift it out vertically. Tilting it can bend the motor shaft.

3. Balance the load

The step that matters most. An uneven load makes the rotor shake, wears the drive and stops the run. Weigh the tubes; do not judge by eye. Opposite positions must match to within 0.1 g, with the same tube type and the same sample in each.

Balanced

Four equal tubes spaced 90° apart.

Runs smoothly

Unbalanced

Two tubes side by side, nothing opposite.

Vibrates, run stops
Sample tubeBlank: water to the same mass

Loading patterns for a 12-place rotor

The tubes must be spread evenly around the axis. When the count does not divide evenly, add a blank tube of water at the same mass.

2 tubesDirectly opposite
3 tubesEvery fourth hole, 120° apart
4 tubes90° apart
5 tubes + 1 blankBlank brings it to 6, 60° apart
6 tubesEvery other hole
7 tubes + 1 blankAll but four holes 90° apart

Fill limits

Fill tubes to no more than 90 % of their nominal volume in swing-out rotors and 80 % in fixed-angle rotors. Close every tube tightly and dry the outside.

The speed limit is the lowest of the rotor, bucket, adapter and tube limits. For commercial tubes, use the speed or × g the tube maker states. Samples denser than usual, such as concentrated salt or sucrose, need a lower speed: ask LORDERAN for the rotor's rated density.

≤ 90 %Swing-out≤ 80 %Fixed-angle

4. Set up the run

  1. 1
    Select the rotorThe number on the screen must match the rotor fitted; the speed limits come from it.
  2. 2
    Enter × g or rpmSwitch Speed / RCF to enter × g directly. The instrument works out the rpm for that rotor. See RPM vs RCF.
  3. 3
    Set the timeBy default the timer starts when the rotor reaches set speed.
  4. 4
    Set the temperature−20 to +40 °C on refrigerated models.
  5. 5
    Set ACC and DECACC 1–10, DEC 0–10. DEC 0 lets the rotor coast to a stop.
  6. 6
    Close the lid and startPress both sides until it clicks. Stay until the rotor reaches set speed and runs quietly.
SIMPLIFIED SCREEN · TDL-5A EXAMPLEROTORNo. 34 × 18 × 7 ml1SPEED / RCF1300 × g≈ 2,550 rpm2TIME00:10:003TEMP20.0 °C4ACC / DEC9 / 95Start6Lid closed and latchedTimer: from set speed

Example: serum gel tubes at 1300 × g for 10 minutes in the TDL-5A 18 × 7 ml rotor (r = 179 mm).

Cold run? Fit the empty rotor and buckets, close the lid and run Pre-cool at a moderate speed, such as 1000 rpm, at your run temperature. Chill large volumes separately in a refrigerator; the chamber cannot cool a full 100 ml bottle quickly.

5. Finish safely

A beep marks the end of the run. The lid stays locked until the rotor has stopped.

  1. 1
    Wait for a full stopThen tap Open and lift the lid with both hands.
  2. 2
    Lift tubes out uprightSo the pellet is not disturbed. Check for spills or broken glass.
  3. 3
    Dry the chamberAfter cold runs, wipe it dry and leave the lid open for a while.
  4. 4
    End of the dayTake the rotor out, store it dry and switch off at the power switch.
Infectious samples, including clinical blood: use sealed tubes in sealed buckets, and load and unload the buckets in a biosafety cabinet. Wait a few minutes after the run before opening the lid. If a tube breaks, keep the lid closed for at least 30 minutes, then decontaminate following your lab's procedure.
Never spin flammable solvents, explosive or peroxide-forming compounds, or highly corrosive samples. These centrifuges are not explosion-proof.
Unusual noise or strong vibration: tap Stop. If the rotor does not slow, switch off at the power switch and wait until it stops before opening the lid.

When an alarm sounds

The screen turns red, the buzzer sounds and the drive stops. Tap the red area to acknowledge, fix the cause, then restart. Note the code and time; alarms are kept in Run Records.

ALARM OR SYMPTOMLIKELY CAUSEWHAT TO DO
Lid not closed (E:3 on the TDL-5A)The lid is not fully latched, or was opened during the run.Press down on both sides until it clicks, then lift lightly to check. If the alarm stays, call service.
OverspeedThe set speed is above the rotor maximum, often because the wrong rotor number is selected.Select the installed rotor and enter a speed within its range. If it appears mid-run, call service.
Rotor identification error (TGL-21M)The rotor number on the screen does not match the rotor fitted.Acknowledge, select the installed rotor, check the settings and start again.
Shaking or loud noiseUsually the load: tubes more than 0.1 g apart, an asymmetric pattern, a missing bucket or a bucket that cannot swing.Tap Stop. Check the load, then the rotor (seated, nut tight, no damage), then the bench. If it persists with a correct load, call service.
Temperature not reachedEnergy-saving mode, blocked vents, a warm room, frequent lid opening or no pre-cool.Check the temperature mode and the 300 mm clearance, pre-cool the rotor and pre-chill large samples.
Lid will not openThe rotor is still turning, or the power failed.Wait for a full stop. After a power failure, unplug, wait until the rotor has stopped and use the emergency release as your manual shows.
Speed sensor faultNo speed signal while the drive runs.Do not restart. Call service.

Cleaning and rotor life

Clean with a soft cloth and a mild neutral detergent. Disinfect with an agent safe for aluminum and paint, such as 70 % ethanol. Do not use chlorine bleach: it corrodes aluminum. Wipe the touch screen with a dry cloth only.

BEFORE EVERY RUNInspect rotor, buckets and tubesCheck the rotor nut is tight and no tube is cracked or deformed.
AFTER EVERY USERemove, clean and dryTake the rotor out, dry the chamber, leave the lid open after cold runs.
WEEKLYEmpty the drip trayDaily if you run cold often.
MONTHLYLook for corrosion and cracksCheck rotors, the lid seal and the space around the vents.
YEARLYClean the condenser finsA service task, or sooner if cooling weakens (TGL-21M schedule).
AFTER 5 YEARSRetire or inspect the rotorLog each rotor’s first-use date; send it for inspection or replace it.

A rotor that has been dropped, or shows cracks, corrosion pits, deformation or damaged hanger slots, stays out of use until LORDERAN has inspected it.

Manuals by model

Each manual has the rotor table, speed limits and alarm list for its model. Choosing a centrifuge? Start with RPM vs RCF: how to convert and choose or the centrifuge collection.

Benchtop, low speedTDL-5A
Benchtop, low speedTDL-8A
Tabletop, large capacityL580R
Benchtop, high speedTGL-21M
Floor, low speedDL-5B / D5M
Floor, low speedDL-5M / D5B
Floor, high speedBH2200R / BH2100R
Floor, high speedBH2500R / BH2300R
Floor, high speedGL-25M / GL-21M
Floor, high speedH2500R / H2100R

Questions

How exactly do the tubes need to be balanced?

The TDL-5A and TGL-21M manuals ask for opposite positions to match within 0.1 g. Weigh the filled tubes or buckets on a balance. Tubes that look equal often differ by more than that.

What if I have an odd number of tubes?

Add a blank tube filled with water to the same mass, so the count fits an even pattern. In a 12-place rotor, three tubes can also run alone if they sit 120° apart, in every fourth hole.

Do empty buckets have to be in the rotor?

Yes. On swing-out rotors every bucket and rack goes on for every run, empty ones included, and opposite buckets must be identical and swing freely.

Can I clean the rotor with bleach?

No. Chlorine-based disinfectants such as sodium hypochlorite corrode aluminum and paint. Use a mild neutral detergent, rinse and dry, and disinfect with 70 % ethanol.

How long does a rotor last?

Five years of service in these manuals. Then return it for inspection or replace it. Write the first-use date on the rotor case so you know when it is due.

Why won't the lid open?

It stays locked until the rotor has stopped. After a power failure, unplug the centrifuge, wait until the rotor is still and use the emergency release as your model's manual describes. Never release the lid while the rotor turns.