How to Use a Thermal Shock Chamber: Install, Load, Program and Maintain

November 21, 2026

Last updated November 2026 · LORDERAN how-to guide

Short answer: a two-zone thermal shock chamber keeps a hot zone and a cold zone at their set points and moves the specimen basket between them. Install it upright, level and with room to breathe, load the basket without blocking the vents, then program each segment with its pre-conditioning value, test temperature and dwell time. Keep the door shut while it runs, stop tests from the controller rather than the power switch, and clean the condenser every three months. Figures come from the LORDERAN two-zone thermal shock chamber user manual.

1. How it works

The hot chamber sits above the cold chamber. A motor with a brake drives the basket between them, and position sensors confirm it has reached the top or bottom. The manual lists three modes: high-temperature testing, low-temperature testing and thermal shock cycling.

Two zones, one basket

The basket carries the specimens up and down.

HotColdBasket moves in about 5 s

One test cycle

High dwell, transfer, low dwell, repeat.

High set pointLow set pointdwelldwelltime → one cycle = high dwell + low dwellRecovery about 5 min
Hot zoneCold zoneBasket and specimen temperature

2. Install it

Installation and electrical connection are for qualified people. The manual says not to run the chamber until it is fully installed and set up.

  1. 1
    Move it uprightKeep it upright in transport and storage so compressor oil stays in place. Use a forklift or lifting straps, never steel cables.
  2. 2
    A level, solid floorOut of sunlight, ventilated, low dust and free of flammable or corrosive gases. Ambient 5 to 35 °C, 45 to 85 %RH. Lower the four feet and level with a spirit level: poor levelling can damage the compressor.
  3. 3
    Leave spaceRoom for the door to open fully and for maintenance at the sides. On air-cooled units, keep more than 0.8 m (31 in) between the condenser exhaust and the wall.
  4. 4
    Power to the nameplateA dedicated breaker rated above the maximum current, a grounded terminal and supply fluctuation under 5 %. Connect with the power off.
  5. 5
    Water and drainWater-cooled units need 5 to 32 °C water with steady flow, a shut-off valve and a 10 bar relief valve; never start with the valves closed. Run the condensate drain downward to a floor drain.

3. Load the basket

The basket moves every cycle, so loose parts and blocked air ducts are the two things to avoid. Airflow sets the temperature uniformity.

  1. 1
    Inside the basketNothing may stick out past the basket. Leave a gap to its edges and fix specimens so they cannot fall or shift when it moves.
  2. 2
    Vents clearNever block the air intake or return vents; blocked airflow increases test error.
  3. 3
    Cables through the portRoute test leads through the test port and seal any gap around the cables.
  4. ✕
    Never insideExplosive, flammable or oxidizing substances: the chamber is not explosion-proof. Corrosive materials such as chlorides and acids shorten the life of the compressor and door gasket. Loose conductive parts can fall into the air-conditioning section.

4. Program and run a test

Keep every value inside the range on your chamber's nameplate. Test values cannot be changed while a test runs.

Write the program

  1. 1
    SegmentsFor each segment set MODE (HIGH or LOW), BSV (pre-heat or pre-cool value), TEST (test-area temperature) and TIME (dwell).
  2. 2
    Time formatWith no unit shown, time is hours.minutes: 1.30 is 1 h 30 min and 0.01 is 1 min.
  3. 3
    CyclesSet the loop count, which zone goes first, and LINK to chain another program.

Set the functions

  1. 1
    WaitingSet the wait bands so the dwell only counts once the zone is in range.
  2. 2
    DefrostBy time or after a number of cold exposures, for example every 5.
  3. 3
    Restart modeSTOP needs a manual restart after a power cut. COLD restarts from segment 1. HOT resumes the interrupted segment.

Run and finish

  1. 1
    RunTap RUN. PAUSE holds the test; STOP ends it. Keep the door shut.
  2. 2
    Stop from the controllerNever cut the power to end a test, and never during high-temperature operation. In COLD or HOT mode the chamber restarts when power returns.
  3. 3
    UnloadLet it return to ambient, open with heat- or cold-resistant gloves, then export the data to USB.

Surfaces above 38 °C or below 0 °C can burn or cause frostbite. Above 55 °C the interior, basket and specimens are very hot. Opening the door with a large temperature difference can damage the equipment, and frequent opening lets moisture in, which can freeze the evaporator.

Alarms and troubleshooting

After fixing the cause, press Reset on the control panel. If the alarm returns, send LORDERAN the model, serial number and alarm. Electrical work is for a qualified electrician with the power off.

ALARM OR SYMPTOMLIKELY CAUSEWHAT TO DO
Hot or cold position faultA position sensor, or the basket did not travel fully; in the cold zone, ice.Check the sensor and clear objects from the basket path. For the cold zone, check for icing.
Door faultA door sensor, or a door opened during the run.Close the door fully and check the sensor for damage or obstruction.
Overtemperature protectionLocal overheating.Stop the test, rearrange the specimens, clear blocked air ducts, remove specimens that generate heat.
Compressor overloadHigh voltage, condenser fan stopped, condenser clogged.Keep 0.8 m behind the exhaust, clean the condenser, check voltage and fan. If the internal protector tripped, wait 6 h before restarting.
Temperature slow to fallSpecimen heat load, iced evaporator, poor cooling.Reduce the heat load, check the condenser, dry the chamber and run a defrost.
Poor uniformityToo many specimens or poor placement.Load fewer specimens, keep them off the vents and adjust the air deflector.

Maintenance

Have the refrigeration, electrical system and compressor checked by a professional once a year. Before a long idle period, dry the chamber: 60 minutes with the door closed, then 15 minutes with it slightly open.

BEFORE EACH TESTSeal and cleanClean the basket, door seals, window and interior. Check the seal with a paper strip around the door.
MONTHLYAlarm checkCheck that the upper and lower deviation alarms act. Back up data and parameters every 2 months.
EVERY 3 MONTHSCondenserBrush or blow dust out of the air-cooled condenser. Never use water. Wear gloves: the fins are sharp.
6 MONTHS AND UPFans, panel, spindleClean the condenser fan and electrical room (vacuum only) every 6 months. Lubricate the spindle every 10,000 cycles.

Manuals and guides

The full manual covers every controller screen, alarm and wiring detail.

User manualThermal shock chamber
Buying guideEnvironmental test chambers
How-to guideClimate chambers
ProductTwo-zone thermal shock chamber

Questions

How fast does the basket move between zones?

The manual lists a 5 second transition and a 5 minute recovery time for the CJ-500. Check your model's specification for its own figures.

Can I test flammable or explosive items?

No. The manual rules out explosive, flammable and oxidizing substances because the chamber is not explosion-proof.

What happens after a power cut?

That depends on the restart mode: STOP waits for you, COLD restarts from the first segment and HOT resumes the segment that was running.

Why does the cold zone ice up?

Moisture enters when the door is opened, especially with a large temperature difference. Keep openings short and set a defrost by time or by number of cold exposures.

How do I get the test data?

Insert a USB drive, select the file and export as XLS or BIN. At the standard 60 s interval the chamber stores about a year of records, then overwrites the oldest.