How to Use a Laminar Flow Hood: Clean Work, First Air and Care
Last updated November 2026 · LORDERAN how-to guide
Short answer: a laminar flow clean bench protects the work, not you: filtered air blows across the work surface and out toward the operator. Use it only for clean, non-hazardous work. Put it out of drafts and traffic, run the fan about 30 minutes before you start, and keep everything downstream of your sample so the first air to reach it comes straight from the HEPA filter. Wipe with alcohol on gauze, never spray the filter, and test the cabinet once a year. Figures come from the LORDERAN CB-H1500/H1800 horizontal laminar flow cabinet manual.
1. Know what it is for
In a horizontal clean bench, HEPA-filtered air comes from the back wall, crosses the work surface and leaves through the open front, toward you. That keeps the work clean to ISO Class 5 but gives you no protection from what you handle.
-
✓Clean, non-hazardous workDust-free assembly of sterile equipment or electronic devices, and other clean work with non-hazardous materials.
-
✕Never for hazardous materialsBMBL says clean benches should never be used when handling cell culture materials, drug formulations, potentially infectious materials or any other potentially hazardous materials. Use a Class II biosafety cabinet for that work.
-
✕No chemicals that give off vaporsThe manual rules out acids, alkalis, and flammable, explosive or volatile hazardous materials. Those belong in a fume hood.
2. Place it
Room air blowing across the opening carries particles onto the work.
-
1Out of drafts and trafficA clean indoor area away from dust and vibration, not in a passage. Keep windows closed and do not let room air blow over the work area or toward the filter.
-
2Room to clean around itIdeally 30 cm (12 in) behind and around the cabinet; at least 8 cm (3 in) at the sides and 3.8 cm (1.5 in) at the back.
-
3Its own grounded circuitA grounded outlet with leakage protection that stays reachable, so the power can be disconnected. Running power is 2,050 W. Assembly is by trained engineers.
-
4Keep the top outlet clearNothing on top of the cabinet, and no water containers on it.
3. Work in first air
The air that reaches your sample should come straight from the filter. Anything between the filter and the sample, including your hands, sheds particles into the airstream that then lands on the work.
Nothing upstream
Sample nearest the filter's airstream; hands work from the front.
Bottle and hand upstream
An object between filter and sample, a hand reaching behind.
4. A work session, step by step
The LORDERAN horizontal cabinets blow 0.2 to 0.45 m/s (40 to 90 fpm) across the work surface through a 99.995 % HEPA filter.
Before
-
1UV, if you use itAt least 30 minutes with nobody in the room. UV only runs with the fan and light off.
-
2Fan for 30 minutesSwitch UV off, the fan on, and let it run about 30 minutes before work.
-
3Wipe downDisinfect the work surface and side walls with alcohol on gauze.
During
-
1Downstream onlyKeep hands, bottles and tools between the sample and you, never between the filter and the sample.
-
2No flames, no paperUse an infrared sterilizer, not a burner. Tissue and other light items can be drawn into the intake.
-
3No vibrationCentrifuges and vortex mixers can shake particles off the filter. Keep loads under 23 kg on any 25 × 25 cm area.
After
-
1Clear and wipeRemove everything and wipe the surfaces dry, then damp, then with alcohol.
-
2UV, if usedAnother 30 minutes with the room empty.
-
3Off and coveredSwitch off and cover the cabinet to keep moisture out.
Troubleshooting
Check the power connection, cable, fuse and switch first. Electrical repairs need a qualified electrician with the power off. If the cabinet behaves abnormally, stop using it and unplug it.
Cleaning and testing
Clean with diluted dishwashing detergent and disinfect with medical alcohol on gauze. Never spray anything at the airflow plate: the HEPA filter must stay dry. Do not spray the control panel or labels.
Manuals and guides
Each LORDERAN laminar flow cabinet has its own manual and datasheet.
Questions
Can I do cell culture in a laminar flow hood?
BMBL says no: clean benches should never be used for cell culture materials, because the air blows toward you. Use a Class II biosafety cabinet.
How long should it run before I start?
The LORDERAN manual asks for about 30 minutes of fan time after switching on, before experiments begin.
Can I spray alcohol inside?
Spray it onto gauze and wipe. Never spray toward the airflow plate, because the HEPA filter has to stay dry.
When does the HEPA filter need replacing?
When the airflow can no longer reach the rated velocity as the filter loads up. Watch the pressure reading, have the airflow measured, and reset the filter timer after a change.
Do I need the UV lamp?
It is optional. If you use it, run it for at least 30 minutes with nobody in the room, check its output regularly and replace the tube after about 600 hours.