Flame Photometer FP6400 Series

Flame Photometer FP6400 Series

Flame photometer is a device used in inorganic chemical analysis to determine the concentration of certain metal ions, K, Na, Li, Ca, Ba.

Repeatability: 1 % CV for 20 consecutive samples
Response Time: <8s
Sample Uptake: <6L/min
Packing Size: 590*540*410mm, 18Kg

    1. LCD screen display.
    2. Direct concentration read-out.
    3. Automatic Calculation of correlation coefficient.
    4. Pre-selection of flame sizes.
    5. Flameout protection device.
    6. Measuring range changing.
    7. LPG burning gas.
    8. Air compressor is included.
  • Mode   FP640 FP6410 FP6430 FP6431 FP6432 FP6440 FP6450
    Sensitivity K 0.01
    (ppm) Na 0.01
      Li - optional 0.01 - - 0.1 0.1
      Ca - - - 2 - 2 2
      Ba - - - - 6 - 6
    Range K 0-100
    (ppm) Na 0-160
      Li - optional 0-100 - - 0-100 0-100
      Ca - - - 0-1000 - 0-1000 0-1000
      Ba - - - - 0-3000 - 0-3000
    Linearity K 0.195 (0.39-3.12)
    (ppm) Na 0.69 (1.15-9.2)
      Li - optional 0.15(0.25-5) - - 0.15 (0.25-5) 0.15 (0.25-5)
      Ca - - - 3 (5-100) - 3 (5-100) 3 (5-100)
      Ba - - - - 9 - 9
    Linearity Error K ≤0.005
    (mmol/L) Na ≤0.03
      Li - - ≤0.021 - - ≤0.021 ≤0.021
      Ca - - - ≤0.075 - ≤0.075 ≤0.075
      Ba - - - - ≤0.066 - ≤0.066
    MDL K ≤0.004
    (mmol/L) Na ≤0.008
      Li - - ≤0.015 - - ≤0.015 ≤0.015
      Ca - - - ≤0.050 - ≤0.050 ≤0.050
      Ba - - - - ≤0.044 - ≤0.044