LORDERAN Climate Chamber -40°C to 150°C, 20-98% RH, Temperature Humidity Test Chamber with Water Recirculation, IEC/ASTM, JS4 Series
- SKU
- JS4010
- Availability
- In stock
- Questions
Cold, dry-heat and damp-heat tests for electronics, automotive parts and materials: −40 °C to +150 °C and 20–98 %RH, in four sizes from 100 to 1000 L.
Hold one condition for days, or program cycles that repeat on their own. Reachable humidity depends on temperature, so check your test point on the chart below before you order.
The picture runs three example set points; colors show the condition, not a camera view.
Performance figures: empty chamber, +25 °C ambient.
Will it reach my humidity?
Every limit is on the humidity chart, taken from the user manual. Drier or cooler points need the low-humidity option in the configurator.
Do I need a water line?
No. A 20 L tank behind the rear cover feeds the humidifier, and condensate drips back into it. One fill lasts up to about 7 days.
Can it run overnight?
Yes. 100 programs of up to 100 segments, alarms on the screen and over the network, and a return to room temperature when the program ends.
Which power does it need?
100 and 250 L: 1-phase 208–240 V, or 3-phase. 500 and 1000 L: 3-phase only. Five 3-phase voltages from 208 to 480 V; all 50/60 Hz.
How it holds your test conditions
The samples sit in a sealed workspace. Everything that changes the air happens in a duct behind its back wall.
Simplified section, not to scale. Part positions follow the user manual; blue arrows show the air path. Screen values are an example.
-
1Load the samples.Space them out on the shelves, keep the load under ¾ of the workspace, and run cables through the Ø100 mm side port.
-
2Set the test.Enter a set point (temperature, humidity, time) on the 7-inch touch screen, or pick a saved program.
-
3Condition the air.A fan pushes air through the duct: the cooling coil chills and dries it, the heater warms it, the humidifier adds steam.
-
4Measure and correct.A dry sensor and a wet-wick sensor read the air; the gap between them gives the humidity. The controller adjusts each part until actual matches set.
-
5End the test.The chamber returns to room temperature to protect the samples. Logged data exports to USB.
Can it reach your humidity at your temperature?
Relative humidity (%RH) compares the water vapor in the air with the most it can hold at that temperature, so the range the chamber can hold changes with temperature. Colored: standard. Hatched: needs the optional low-humidity function.
Specifications
Performance figures: empty chamber, +25 °C ambient, clean condenser. Fluctuation is how far a reading swings around the set point once stable; deviation and uniformity compare points across the workspace.
Differs between models
| Model | 100 L | 250 L | 500 L | 1000 L |
|---|---|---|---|---|
| SKU | JS4010 | JS4025 | JS4050 | JS4100 |
| Interior, W × D × H | 450 × 450 × 500 mm | 600 × 600 × 700 mm | 800 × 700 × 900 mm | 1000 × 1000 × 1000 mm |
| Exterior, W × D × H, casters included | 780 × 1130 × 1580 mm | 960 × 1500 × 1760 mm | 1130 × 1550 × 1950 mm | 1320 × 1890 × 2050 mm |
| Shelves (standard) | 2 | 2 | 3 | 3 |
| Weight (approx.) | 250 kg | 320 kg | 500 kg | 680 kg |
| Noise level | 60–65 dB | 60–65 dB | 65–70 dB | 65–70 dB |
| Power supply (50/60 Hz) | 1-phase: 208–240 V 3-phase: 208, 220–240, 400, 415 or 480 V |
1-phase: 208–240 V 3-phase: 208, 220–240, 400, 415 or 480 V |
3-phase: 208, 220–240, 400, 415 or 480 V | 3-phase: 208, 220–240, 400, 415 or 480 V |
| Rated power | 6 kW | 7 kW | 10 kW | 13 kW |
| Rated current | 27 A at 220 V, 1-phase | 32 A at 220 V, 1-phase | 27 A at 220 V 12 A at 480 V |
34 A at 220 V 16 A at 480 V |
| Temperature range | −40 to +150 °C |
|---|---|
| Humidity range | 20–98 %RH (see chart) |
| Temperature fluctuation | ±0.5 °C |
| Temperature deviation | ≤ ±2 °C |
| Temperature uniformity | ≤ 2 °C |
| Humidity fluctuation | ±2–3 %RH |
| Humidity deviation | ±2.0 %RH above 75 %RH ±3.0 %RH at 75 %RH or below |
| Humidity uniformity | ≤ 5 %RH |
| Heating rate (average) | 2–3 °C/min |
| Cooling rate (average) | 0.7–1 °C/min |
| Heating-up time | +20 °C to +150 °C in ≤ 60 min |
| Cooling-down time | +20 °C to −40 °C in ≤ 70 min |
| Refrigeration | Air-cooled, R449A |
|---|---|
| Controller | 7-inch color touch screen, 100 programs × 100 segments |
| Interfaces | USB (data and program export), RS-485 (PC monitoring and control) |
| Inner chamber and shell | SUS304 stainless steel: 2 mm fully welded inner chamber, 1.5 mm outer shell |
| Insulation | PU foam and rock wool |
| Door | Double silicone seals, heated double-glazed window, interior light |
| Test port | Ø100 mm with rubber plug, left side (custom ports on request) |
| Water tank | 20 L, built in; RO or DI water |
Which size fits your samples and your doorways?
Keep the samples to about ¾ of the workspace, with gaps between items and the walls. The drawing puts each size beside a standard 36 × 80 in door: the chamber travels upright, so check its narrowest side and its height against every doorway on the route.
Exterior dimensions as published. Height includes the casters; packaging not included.
100 L
100 L · JS4010
- Exterior
- 780 × 1130 × 1580 mmW × D × H, casters included
- Interior
- 450 × 450 × 500 mm2 shelves as standard
- Samples
- up to about 75 L in total¾ of the workspace, with gaps
- Weight
- 250 kg
- Power
- 1-phase: 208–240 V
3-phase: 208, 220–240, 400, 415 or 480 V50/60 Hz
250 L
250 L · JS4025
- Exterior
- 960 × 1500 × 1760 mmW × D × H, casters included
- Interior
- 600 × 600 × 700 mm2 shelves as standard
- Samples
- up to about 190 L in total¾ of the workspace, with gaps
- Weight
- 320 kg
- Power
- 1-phase: 208–240 V
3-phase: 208, 220–240, 400, 415 or 480 V50/60 Hz
500 L
500 L · JS4050
- Exterior
- 1130 × 1550 × 1950 mmW × D × H, casters included
- Interior
- 800 × 700 × 900 mm3 shelves as standard
- Samples
- up to about 380 L in total¾ of the workspace, with gaps
- Weight
- 500 kg
- Power
- 3-phase: 208, 220–240, 400, 415 or 480 V50/60 Hz
1000 L
1000 L · JS4100
- Exterior
- 1320 × 1890 × 2050 mmW × D × H, casters included
- Interior
- 1000 × 1000 × 1000 mm3 shelves as standard
- Samples
- up to about 750 L in total¾ of the workspace, with gaps
- Weight
- 680 kg
- Power
- 3-phase: 208, 220–240, 400, 415 or 480 V50/60 Hz
Tap a size for its full dimensions, weight and power.
What to prepare before delivery
Power, room, water and access. The user manual has the details.
Dedicated circuit with breaker and RCD sized to the rated current; earth ≤ 4 Ω.
+5 to +32 °C, ≤ 85 %RH, ventilated, out of direct sun.
Level and rated for the chamber, 250 to 680 kg.
40 to 80 cm around the chamber, plus room to open the door fully.
RO or DI water only, ≤ 5 µS/cm. No tap water.
A floor drain, or a tray under the rear overflow.
Duty-paid to your loading dock. Plan a forklift or pallet jack.
Shelves, USB stick, spare parts, door key and documents.
Which test profiles can you run?
A test is a list of set points: what to reach and how long to hold it. These are the three common kinds.
Temperature, left axisHumidity, right axis
Set-point examples drawn for explanation, not measured chamber data. A program can’t ramp faster than the chamber: 2–3 °C/min up and 0.7–1.0 °C/min down on average, empty chamber at +25 °C ambient. Heavy, wet or heat-producing loads are slower.
What keeps a long test running
Named components, the build, and the protections behind an unattended test.
Overheat and over-pressure protection, vibration damping. R449A, air-cooled: no cooling water.
100 programs × 100 segments, loops and a wait step. Data logs every 60 s and exports to USB as XLS.
Live curves, compressor status and alarms over the network. RS-485 for PC control.
Air circulation keeps the ceiling dry, so condensate doesn’t drip onto samples.
Back to room temperature when a program ends, to protect the samples.
Double silicone seals, rated for more than 1000 hours of frost-free running.
2 mm fully welded inner chamber, 1.5 mm shell, PU foam and rock wool.
Double glazing with an interior light. Ø100 mm port on the left; custom ports on request.
Show all safety protections
- Sample over-temperature protection (controller)
- Independent over-temperature protector
- Compressor overload protection
- Compressor over-pressure protection
- Fan motor overload protection
- Humidifier dry-burn protection
- Water shortage protection
- Phase loss and reversal protection
- Short-circuit and earth-leakage protection
- Three-color alarm light
- Flame-retardant non-metallic materials
Show the reference standards: IEC 60068, GB/T, GJB150
Need colder, faster, or a thermal shock?
Pick by the stress your product has to survive: a steady climate, fast ramps, or a sudden hot-to-cold transfer.
Questions before you order
Options, water, delivery, warranty, returns and quotes.
Do I need the low-humidity function?
Only if a test point falls in the hatched zone of the humidity chart: below 20 %RH (down to 10 %RH from 40 °C), below the standard lower limit between 20 °C and 45 °C, or any humidity test from 10 °C to 20 °C. Tell us your test temperature when you order and we tune the system before shipping.
Do I need a water connection?
No. Steam comes from purified water in a 20 L tank behind the rear cover. Water the cooling coil pulls out of the air drips back into the tank by gravity, so most of it is used again.
- Tank
- 20 L, built in
- One fill
- up to about 7 days in normal use
- Water
- RO or DI only, ≤ 5 µS/cm; never tap water
- Reuse
- over 80% higher water utilization*
*Manufacturer figure, compared with a chamber that drains its condensate. Long high-humidity tests use more water. Automatic top-up from a purified-water supply is available on request. Monthly: change the water and clean the tank.
How is it delivered, and when does it ship?
Freight, duties and taxes are paid to your loading dock; moving the chamber inside is not included. Your confirmed ship date is included in the order acknowledgement.
What does the warranty cover?
A 3-year worldwide warranty from the date of purchase, covering core components and controls. We repair first; if a major defect can’t be repaired, we replace the unit or give a full refund. Seals, heating tubes and other wear parts are excluded. A 2-year extended warranty is available.
Can I return it?
Yes, within 30 days of delivery if unused and in original condition. For non-quality returns the buyer pays return shipping, a 3% restocking fee and the card processing fee. Defective or damaged items are covered in full.
Can I get a formal quote for a purchase order?
Yes. Select your size and options, then click Get a quote for a PDF quotation with DDP shipping included.
Choose size, voltage and options
The price, delivery date and PDF quote update with your choices.