LORDERAN Horizontal Laminar Flow Cabinet with HEPA Filter, ISO Class 5 Clean Bench for Laboratory & PCR Cell Culture
- SKU
- H1100-110
- Availability
- In stock
- Questions
A horizontal laminar flow clean bench for sterile, non-hazardous work: HEPA-filtered air crosses the work surface from the back of the cabinet toward you, so particles from the operator and the room are carried away from the sample, giving an ISO Class 5 work zone. Four widths, open-front or with a motorized window, supplied with a base stand. It does not protect the operator.
Does it protect the operator?
No. It protects the sample by blowing clean air toward you. For infectious or hazardous materials, choose a Class II biosafety cabinet.
Open front or window?
H1100 and H1500 are open-front. H1300 and H1800 add a motorized front window opening to 430 mm.
When does the filter need changing?
The HEPA filter is rated for 5,000 hours; the controller raises a replacement alarm at 3,500 hours.
Which power does it need?
110 V or 220 V for every model. Choose the voltage in the configurator.
Clean air that flows toward you
A fan draws room air through the washable pre-filter and pushes it through a HEPA filter covering the back wall. The filtered air crosses the work surface in parallel, unidirectional streams and leaves through the open front, so the sample always sits upstream of your hands, your breath and the room.
Side section. The cabinet is held under constant positive pressure, so room air cannot drift in.
- Glass-fibre HEPA filter, damp-proof and fire-proof.
- Washable polyester-fibre pre-filter protects the HEPA filter.
- Laminar flow cabinets protect the sample only. For infectious or hazardous materials, choose a biosafety cabinet.
- Transparent side glass brightens the work zone.
What ISO Class 5 means
ISO 14644-1 sets the most particles of 0.5 μm and larger that a cubic meter of air may hold: 3,520 for ISO Class 5. Per liter, the volume of a large bottle, that is about three or four particles; each class above allows ten times more.
Limits from ISO 14644-1 for particles ≥ 0.5 μm, shown per liter (1/1000 m³). Ordinary indoor air is typically ISO 8 to 9, up to 3,520 to 35,200 particles per liter. Class 100 is the older US FED-STD-209E name for ISO 5.
What a clean bench protects
A laminar flow cabinet blows clean air over the work and out toward you, so it keeps the sample clean but sends anything the sample releases into your face and the room. For infectious material, choose a Class II biosafety cabinet; for chemical fumes, a fume hood.
Specifications
Models H1100, H1500, H1300 and H1800, each available in 110 V and 220 V versions. "/" means not applicable to the open-front models.
Differs between models
| Model | H1100 | H1500 | H1300 | H1800 |
|---|---|---|---|---|
| Dimensions | ||||
| External Size (mm) (w*d*h) | 1100x808x1690 | 1500x808x1690 | 1300x825x2040 | 1800x820x2070 |
| Internal Size (mm) (w*d*h) | 1000x500x600 | 1400x500x600 | 1200x500x570 | 1700x500x570 |
| Work Surface Height | 750 mm | |||
| Max Opening | / | / | 430 mm | 430 mm |
| Gross Weight | 175 kg | 216 kg | 228 kg | 306 kg |
| Packing Size (mm) | 1250x1050x1350 | 1650x1050x1350 | 1450x1060x1650 | 1950x1030x1540 |
| Performance | ||||
| Air Velocity | 0.3 to 0.45 m/s | 0.3 to 0.45 m/s | 0.3 to 0.5 m/s | 0.3 to 0.5 m/s |
| Noise | <65 dB | |||
| Cabinet | ||||
| LED Lamp | 12 W*1 | 16 W*1 | 12 W*1 | 16 W*1 |
| UV Light | Yes | |||
| Front Window | / | / | Motorized | Motorized |
| Consumption | 200 W | 300 W | 400 W | 450 W |
Horizontal or vertical: how to choose
Horizontal hoods push air across the work surface and give the strongest protection to the work itself. Vertical hoods push air down, which suits large equipment that would block a horizontal stream.
Two families, four widths
Open-front H1100 and H1500 for easy access and fast work. H1300 and H1800 add a motorized front window and a taller cabinet. Front elevations drawn to scale.
H1100 · H1500
Open front for unobstructed reach. 1000 and 1400 mm internal width, 0.3–0.45 m/s airflow.
H1300 · H1800
Motorized front window opening to 430 mm. 1200 and 1700 mm internal width, 0.3–0.5 m/s airflow.
A filter you can plan around
The HEPA filter is rated for 5,000 hours. The controller counts system work time and raises a replacement alarm at 3,500 hours, so the filter can be changed on schedule, well before performance drops.
Glass-fibre HEPA filter, 99.999% or better for particles of 0.3 μm and larger.
Details that make daily work easier
A clear display, safe UV handling and practical touches for a bench that is used every day.
Controls and safety
The LCD shows airflow velocity, UV timer, UV work time, system work time and real time. A UV start-up delay protects the operator, and the lamps are interlocked: switching on the fluorescent lamp turns the UV lamp off.
- UV253.7 nm, the most efficient decontamination wavelength
- LEDrecessed lighting to reduce eye fatigue
- Socketstwo waterproof sockets on the side panels
- Installassemble the base, then fix the main body
Everything at a glance
Every model shares the same filtration, lighting and controls.
Show all 15 features: Clean air · Controls · Build
Clean air
- HEPA filter, 99.999% at ≥0.3 μm
- Damp-proof, fire-proof glass-fibre media
- Washable polyester-fibre pre-filter
- ISO Class 5 work zone
- Constant positive pressure
Controls
- LCD with airflow velocity and timers
- UV timer and UV work-time record
- System work-time record and real-time clock
- UV start-up delay
- Fluorescent and UV lamp interlock
Build
- Transparent side glass windows
- Recessed LED lighting
- Two waterproof side sockets
- Base stand included
- 3-year worldwide warranty
Download brochures
The full catalog and the user manuals for each model family, as PDF files.
Questions buyers ask
Warranty, returns, delivery and quotes.
What does the warranty cover?
A 3-year worldwide warranty from the date of purchase, covering core components and controls. We repair first; if a major defect can’t be repaired, we replace the unit or give a full refund. HEPA filters, seals and other wear parts are excluded.
Can I return it?
Yes, within 30 days of delivery if unused and in original condition. For non-quality returns the buyer pays return shipping, a 3% restocking fee and the card processing fee. Defective or damaged items are covered in full.
How is it delivered, and when does it ship?
Freight, duties and taxes are paid to your loading dock; moving the cabinet inside is not included. Your confirmed ship date is included in the order acknowledgement.
Can I get a formal quote for a purchase order?
Yes. Select your size and options, then click Get a quote for a PDF quotation with DDP shipping included.
Choose model and voltage