How to Use a Climate Chamber: Setup, Water, Programs and Care

November 21, 2026

Last updated November 2026 · LORDERAN how-to guide

Short answer: keep the chamber upright, give it a level floor, a dedicated grounded circuit and room around it, and fill the built-in tank with RO or DI water. Check that your test point sits inside the humidity range, space the samples so air can move, and set the test as one fixed condition or as a program of segments and loops. After cold or humid tests, dry the chamber at 60 °C before you switch off. The figures come from the LORDERAN JS4 climate chamber manual; the JS6 uses the same built-in water system.

1. Install it

  1. 1
    Keep it uprightMove it upright on textile slings, never steel cables. If it was tilted, or moved in very cold weather, let it stand upright at room temperature for at least 24 hours before the refrigeration starts.
  2. 2
    Level, stable floorCheck with a spirit level front to back and side to side. The 100 L stands on braked casters; the larger sizes have adjustable feet.
  3. 3
    The right room5 to 32 °C (41 to 90 °F) and no more than 85 %RH, out of direct sun and away from flammable materials. No fire detector directly above.
  4. 4
    Room around itClearances of 40, 60 and 80 cm (16, 24 and 31 in), with space to open the door fully and for condenser air and service at the back.
  5. 5
    Its own circuitA dedicated supply with a breaker and residual-current protection, ground resistance of 4 Ω or less, connected by a qualified electrician. Never use neutral as ground.
ChamberRear: condenser air outFront: dooropens fullySide:serviceSide:serviceMinimum clearances 40, 60 and 80 cm: positions in Figure 2-4 of the manual

Measure the route in. The 100 L is 780 mm (30.7 in) wide and fits a standard 36 × 80 in door; the 250, 500 and 1000 L are wider than that door.

2. Fill the water

The humidifier draws from a 20 L tank behind the rear cover, and the water that condenses on the cooling coil runs back into it. There is no water line to connect. One fill lasts up to about 7 days in normal use.

WorkspaceAir ductCooling coilHeaterHumidifierTank, 20 LRO or DI waterpumpCondensate runs backby gravityOverflow to tray
Air pathWaterHeat

Which water

  1. 1
    RO or DI waterConductivity of 5 µS/cm or less when you fill. Distilled water is acceptable.
  2. 2
    Never tap waterNor well, mineral or softened water, or water with additives.
  3. 3
    Replace it monthlyMeasure the conductivity each month and change the water at least that often, sooner if it turns cloudy or colored.

First fill

  1. 1
    Fill to ½–¾Open the rear cover with the triangular key and fill through the filling port. Keep it at least half full, even for temperature-only tests.
  2. 2
    Prime the systemSwitch on and allow a few minutes. If DI3 or DI8 appears, squeeze the tube between the level switch and the humidifier to push out the air.
  3. 3
    Run a short testFor example 40 °C at 90 %RH for 30 minutes.
  4. 4
    Check the overflowSurplus water leaves under the rear cover into a tray. Keep the pipe end open to air, not under water.

3. Check your test point

Relative humidity depends on temperature, so the humidity a chamber can hold changes with the set temperature. Check the point before you program it.

  1. ✓
    Standard range98 %RH from 20 to 70 °C, easing to 90 %RH at 90 °C. The lower limit is 20 %RH from 45 °C up, rising to 55 %RH at 20 °C.
  2. +
    Low-humidity optionNeeded for any humidity test between 10 and 20 °C, and for down to 10 %RH from 40 °C. It adds a compressed-air dryer.
  3. −
    Outside 10 to 90 °CRun temperature only, with humidity set to OFF, or ask LORDERAN.
010203040506070809010002040608010040 °C · 93 %RH ✓85 °C · 85 %RH ✓25 °C · 30 %RH: optionTemperature (°C)Relative humidity (%RH)

Blue: standard zone. Hatched: needs the low-humidity option. JS4 manual, section 3.2.

Back-wall ductDoor✕✕Not on the floorNot at the outlet

4. Load the samples

  1. 1
    Under ¾ of the workspaceSpread the samples out and stagger them so air reaches every side. Stand large flat samples on edge, in the direction of the airflow.
  2. 2
    Keep the air paths clearNothing on the workspace floor and nothing in front of the air outlet in the back wall.
  3. 3
    Cables through the portThe Ø100 mm (3.9 in) test port is on the left side. Plug it when unused and seal around any cables.
  4. 4
    Powered samplesTell LORDERAN in advance if they give off more than a few hundred watts of heat.
Never put in explosive, flammable, oxidizing or corrosive substances, batteries or products that may vent or burn, conductive dust, food, medicine, biological material or living things.

5. Set up the test

Close the supply breaker, press POWER and log in; change the default password the first time. The chamber starts in STOP. Choose FIX for one condition or PROG for a profile.

  1. 1
    FIX: one conditionSet temperature and humidity, or humidity OFF for a temperature-only test. RUN TIME is continuous or a set duration. TEMP SLOPE and HUMI SLOPE limit how fast the set point moves; 0 means no limit.
  2. 2
    PROG: segmentsUp to 100 programs of 100 segments. Each segment has a target temperature, a target humidity or OFF, and a time in h.mm, so 1.30 is 1 h 30 min. The set point moves to the new target within the segment time.
  3. 3
    LoopsBEGIN and END mark the segments to repeat and CYCLE says how many times they run in total, with up to four loops per program. PROG CYCLE repeats the whole program (0 means endless) and LINK PROG NO. starts another one after it.
  4. 4
    WAITWith WAIT on, a segment does not end until the chamber is inside the wait zone or the maximum wait time has passed. Use it so a hold does not start before the chamber gets there.
  5. 5
    BOOT MODEDecide what happens after a power failure: STOP stays stopped, COLD restarts the program, HOT resumes where it stopped.

A running program cannot be edited. Export programs to USB first as a backup.

-40025850 h2 h4 h6 h8 h10 h12 h1234Loop: segments 1–4× 3 in all°C
SEGMENT
TEMP
HUMI
TIME (h.mm)
01
−40.0
OFF
1.10
02
−40.0
OFF
1.00
03
85.0
OFF
1.10
04
85.0
OFF
1.00

Loop BEGIN 01, END 04, CYCLE 3: the loop runs three times in total.

6. Finish and shut down

When a program ends, "Finished" appears and the chamber returns to room temperature.

  1. 1
    Check the temperature firstAbove 55 °C the escaping air is dangerous: wear heat-resistant gloves. Below 0 °C wear cold-resistant gloves and never touch frozen metal with bare or wet skin.
  2. 2
    Dry after cold testsAfter a sub-zero test, run 60 °C for 30 minutes before switching off.
  3. 3
    Dry after humid testsRun 60 °C with humidity OFF for about an hour, then wipe water from the workspace.
  4. 4
    Switch off in orderPress STOP, confirm, remove the samples, then press POWER. For longer breaks, also switch off the breaker.
Never cut the main power during a hot run. Stop the test on the screen first.
Keep the door closed while it runs, and do not touch surfaces above +38 °C or below 0 °C.
The condenser fan can start at any time while the chamber has power.
Ice on the cooling coil: open the door and defrost by running at a high temperature, such as 60 °C. Below 0 °C the wet-bulb gauze freezes too, so allow a few minutes above 0 °C before a humid phase.

Alarms

A red background means the fault is still there and CLEAR is disabled. Once the background turns green, touch CLEAR to reset.

ALARMCAUSEWHAT TO DO
DI1 · Over-temperature (controller)The chamber went past the alarm limit. The chamber switches off.Find the cause, let it cool, then CLEAR.
DI2 · Over-temperature (knob protector)The independent protector tripped. The chamber switches off.Let it cool. Set the knob about 15 °C above your highest test temperature.
DI3 · Humidifier dry-burnToo little water at the humidifier: low tank, air in the hose, pump or level switch.Top up, squeeze the tube between level switch and humidifier to clear air, then CLEAR.
DI4 / DI5 · Compressor overloadThe compressor overheated or was overloaded. The chamber cannot start.Clean the condenser, check supply voltage and room temperature. An electrician resets the relay.
DI6 · Phase sequenceThree-phase supply wired in the wrong order. The chamber cannot start.An electrician swaps two phases.
DI8 · Low waterThe tank is below the level switch. Humidifying stops.Top up, wait for the green background, then CLEAR.
DI10 · Circulation fanThe fan motor protection tripped.Have the fan motor checked, then CLEAR.
Screen dark, breaker trippedShort circuit or leakage current.An electrician checks the supply and resets once. If it trips again, leave it off and contact LORDERAN.

Maintenance

Clean the workspace with a neutral detergent and clean water, never abrasives, steel wool, chlorine cleaners or solvents. Never use water or spray on the condenser. Use only original LORDERAN gauze, wetted with purified water.

BEFORE EACH TESTWater, tray and wickCheck the tank level, the overflow tray, the wet-bulb gauze and that the workspace is clean.
AFTER A HUMID TESTDry it outRun 60 °C with humidity OFF for about 1 hour, then wipe out any water.
WEEKLYHoses and trayCheck the hose joints and empty the overflow tray.
MONTHLYWater and sealsMeasure the water conductivity and replace the water. Check the door seals.
EVERY 3 MONTHSCondenser and gauzeBrush or blow the condenser clean against the airflow. Fit new wet-bulb gauze.
EVERY 6 MONTHSElectrical and scaleAn electrician cleans the electrical compartment and tests both over-temperature protectors; the humidifier is checked for scale.
YEARLYCalibrationCalibration by an accredited lab and a refrigeration service.
BREAK OVER 2 WEEKSDrain itRun 60 °C with humidity OFF for 30 min, switch off, empty and dry the tank, leave the door ajar.

Manuals and models

The JS4 manual covers installation, programming, water and maintenance in full. Choosing a chamber? Start with how to choose an environmental test chamber.

−40 to +150 °C · 100–1000 L · from $9,999JS4 climate chamber
−70 to +150 °C · 100–1000 L · from $12,999JS6 climate chamber
All instrumentsManuals and datasheets
All test chambersCollection

Questions

Can I fill it with tap water?

No. The manual calls for RO or DI water at 5 µS/cm or less, and accepts distilled water. Tap, well, mineral and softened water are ruled out, as is water with additives.

How often do I refill the tank?

One fill lasts up to about 7 days in normal use, because some water leaves as vapor when the door opens and through the overflow. Check the level before each humidity test and keep it at least half full.

Do I need a water line or a drain?

No water line: the 20 L tank is built in. Surplus water leaves through the overflow into a tray, which you empty weekly.

Why can't it hold 30 %RH at 25 °C?

That point is below the standard lower limit, which rises toward 55 %RH at 20 °C. It needs the optional low-humidity function, which also allows down to 10 %RH from 40 °C.

What happens if the power fails mid-test?

That depends on BOOT MODE: STOP stays stopped, COLD restarts the program from the beginning, and HOT resumes from the interrupted point. Set it before a long test.

How do I stop ice and corrosion building up?

After a sub-zero test run 60 °C for 30 minutes, and after a humidity test run 60 °C with humidity OFF for about an hour, then wipe out any water before switching off.