Environmental test chamber

LORDERAN Rapid Temperature Change Chamber -70°C to 150°C, 5/10/15°C/min Thermal Cycling Test Chamber for Semiconductor & Auto

SKU
KB6010
Availability
In stock
Questions

Ask about this instrument

LORDERAN Rapid Temperature Change Chamber -70°C to 150°C, 5/10/15°C/min Thermal Cycling Test Chamber for Semiconductor & Auto

An engineer replies by email or phone, usually within 12 hours.

Capacity · 100L
Heating/Cooling Rate (Non-linear) · 5°C/min

Environmental stress screening and temperature cycling: −70 to 150 °C at an average 5, 10 or 15 °C/min, with 20 to 98 %RH, in four sizes from 100 to 1000 L. Fast swings expose weak solder joints, cracked packages and marginal components before products ship; which rates you can order, and whether they need a water chiller, depends on the size.

Example cycle+125 °Chot dwell15 °C/minramping down−55 °Ccold dwell15 °C/minramping up+125−55Hot dwellRamp, 15 °C/minCold dwellExample cycle+125 °Chot dwell15 °C/minramping down−55 °Ccold dwell15 °C/minramping upHotRampCold
−70 → 150 °Ctemperature range
20–98 %RHhumidity range
5 · 10 · 15°C/min ramp rates
100–1000 Lfour chamber sizes
3 yearswarranty
SemiconductorsAutomotive electronicsPCB assembliesAerospace and defenseBatteries and power modules

How fast is 15 °C/min?

About 15 min for the full −70 to +150 °C sweep; 10 °C/min takes about 22 min, 5 °C/min about 44 min. Average rates, no load.

Air or water cooling?

Most versions let you choose. Water only: 250 and 500 L at 15 °C/min, 1000 L at 10 or 15 °C/min, with a 12 to 25 HP chiller.

Do I need a water line?

No. The humidifier runs from a built-in tank of RO or DI water, and condensate drains back into it by gravity.

Which power does it need?

Three-phase: 208, 220–240, 400, 415 or 480 V, 50/60 Hz. Rated power 8 to 65.5 kW, by size, rate and cooling type.

Rapid Temperature Change Test Chamber documentsAll four sizes, ramp rates, power, dimensions and test standards in one document to share with your team. The user manual covers installation, programming, safety and maintenance.Download datasheet Download user manual

Choose how fast the temperature moves

The ramp rate sets how hard each cycle stresses the product. At 15 °C per minute the chamber crosses the full 220 °C span in about a quarter of an hour, so more cycles fit into every test day and failures surface sooner.

−70 °C−40 °C0 °C25 °C85 °C150 °C01020304050min15 °C/min≈ 15 min10 °C/min≈ 22 min5 °C/min≈ 44 min

Rates are non-linear averages measured with no load. A heavy or high-mass sample slows the ramp.

±0.5 °Ctemperature fluctuation
±1.0 °Ctemperature deviation
±2–3 %RHhumidity fluctuation
7.0 intouch screen
  • Mechanical binary (cascade) refrigeration reaches −70 °C.
  • Ceramic heating elements with PID control for tight, even heating.
  • Broadband forced air circulation, out at the top and in at the bottom.
  • Set different variable rates for different stress-screening profiles.

What a faster rate buys you

Screening profiles alternate hot and cold dwells, and every ramp between them is test time. In the same two hours, the 15 °C/min version completes 3 full −40 ↔ +85 °C cycles and the 5 °C/min version completes 1, so a campaign of a few hundred cycles finishes in a fraction of the days.

+85 °C+25 °C0 °C−40 °C0306090120minutes123115 °C/min3 full cyclesin 2 hours5 °C/min1 full cyclein 2 hours+85 °C+25 °C0 °C−40 °C0306090120minutes123115 °C/min: 3 full cycles in 2 h5 °C/min: 1 full cycle in 2 h

Example set-point profile: −40 ↔ +85 °C with 10 min dwells, drawn from the published average rates, no load; a heavy sample slows every ramp. Not measured chamber data.

How it moves the temperature that fast

Strong heating and a two-stage refrigeration system drive the change; forced air carries it to every shelf.

WorkspaceAir ductCable portSamples on shelvesComp.Cascade refrigeration, air- or water-cooledFanCooling coilCeramic heaterHumidifier, makes steamWater tankTemperature and humidity sensorsTouch screenSet40.0 °C93 %RH12345Touch screenSet40.0 °C93 %RHWorkspaceAir duct, behind back wallCable portSamples on shelvesFanCooling coilHeaterHumidifierComp.Refrigeration unitcascade, air or waterWater tankTemperature/RH sensors12345

Simplified section, not to scale. Blue arrows show the air path: out at the top, back at the bottom. Screen values are an example.

  1. 1
    Load the samples.Space them out on the shelves and keep the load light: mass slows every ramp. Cables go through the 100 mm test hole.
  2. 2
    Set the profile.On the 7-inch touch screen, set temperature, humidity and time for each step, with its own rate. 100 programs × 100 steps.
  3. 3
    Ramp.Ceramic heaters and the cascade refrigeration change the air temperature; the fan sends the air out at the top and draws it back at the bottom.
  4. 4
    Measure and correct.PID control compares actual with set values and trims heating, cooling and humidity together.
  5. 5
    Record.The test curve is logged with no time limit. Alarms and live data reach your PC or phone.

Specifications

Four chamber sizes, each available at 5, 10 or 15 °C/min.

Download the full datasheet (PDF)

Differs between models

Model 100 L 250 L 500 L 1000 L
Model and size
SKU KB6010 KB6025 KB6050 KB6100
Interior Size (W × D × H) 450 × 450 × 500 mm 600 × 600 × 700 mm 800 × 700 × 900 mm 1000 × 1000 × 1000 mm
Shelves 2 2 3 3
Exterior and packing size (W × D × H)
Exterior · 5 °C/min water-cooled 780 × 1300 × 1580 mm 960 × 1500 × 1760 mm 1130 × 1550 × 1950 mm 1320 × 1890 × 2050 mm
Exterior · 5 °C/min air-cooled, 10/15 °C/min 780 × 1600 × 1700 mm 960 × 2420 × 1760 mm 1130 × 2500 × 1980 mm 1320 × 2800 × 2080 mm
Packing · 5 °C/min water-cooled 850 × 1450 × 1900 mm 1100 × 1610 × 1980 mm 1250 × 1650 × 2160 mm 1450 × 2050 × 2250 mm
Packing · 5 °C/min air-cooled, 10/15 °C/min 930 × 1860 × 1840 mm 1100 × 2550 × 1910 mm 1300 × 2700 × 2200 mm 1450 × 2960 × 2300 mm
Packing note Outer dimensions of the export crate, pallet included
Performance
Temperature range −70 to 150 °C
Humidity range 20 to 98 %RH
Temperature fluctuation ±0.5 °C
Humidity fluctuation ±2–3 %RH
Temperature deviation ±1.0 °C
Humidity deviation ±2.0 %RH (>75%); ±3.0 %RH (≤ 75%)
Heating-up time Non-linear average 5/10/15 °C/min, no load.
Cooling down time Non-linear average 5/10/15 °C/min, no load.
Control method Balanced temperature and humidity control (BTHC) with PID intelligent adjustment
Power and cooling
Power, 5 °C/min 8 kW (air-cooled) 10 kW (air-cooled) 13 kW (air-cooled) 28.5 kW (water-cooled)
Power, 10 °C/min 11 kW (air-cooled) 23 kW (air-cooled) 29 kW (water-cooled) 47 kW (water-cooled)
Power, 15 °C/min 13.5 kW (air-cooled) 25 kW (water-cooled) 44 kW (water-cooled) 65.5 kW (water-cooled)
Cooling, 5 °C/min Air, or water with a 5 HP chiller Air, or water with a 5 HP chiller Air, or water with a 8 HP chiller Air, or water with a 12 HP chiller
Cooling, 10 °C/min Air, or water with a 8 HP chiller Air, or water with a 12 HP chiller Air, or water with a 12 HP chiller Water only, 20 HP chiller
Cooling, 15 °C/min Air, or water with a 8 HP chiller Water only, 12 HP chiller Water only, 15 HP chiller Water only, 25 HP chiller
Refrigeration system Mechanical binary (cascade) refrigeration
Power supply 3-phase: 208, 220–240, 400, 415 or 480 V; 50/60 Hz
Construction
Display 7-inch touch screen
Inner Body Material SUS304 mirror stainless steel plate
Insulation Material Special ultra-fine glass wool + polyurethane foam for rapid temperature change test chamber
Airflow Mode Broadband forced air circulation (top out and bottom in)
Humidifier Device Special stainless-steel humidifier

Will it fit your lab?

Check doorways and floor space before you order. Front views are drawn to scale beside a standard 36 × 80 in door.

Standard door36 × 80 in914 × 2032 mm100L · KB6010W 780 mmD 1600 · H 1700Fits through the door250L · KB6025W 960 mmD 2420 · H 1760Wider than door500L · KB6050W 1130 mmD 2500 · H 1980Wider than door1000L · KB6100W 1320 mmD 2800 · H 2080Wider and taller than door80 in door heightDoor36 × 80 in100 L✓ fits250 L✗ wider500 L✗ wider1000 L✗ wider✗ taller

Drawn at the largest configuration. The 5 °C/min water-cooled models are shallower: D 1300 / 1500 / 1550 / 1890 mm. Measure the clear opening of your door, which is smaller than the door leaf. Crate sizes are in the specification table.

Size, speed and cooling at a glance

Faster ramps and larger chambers need more refrigeration. Each cell shows the power rating and the cooling you can order for that size and rate. Water cooling runs on an external chiller of the size shown.

Chamber / rate
5 °C/min
10 °C/min
15 °C/min
100L
8 kWrated air-cooledair or water · 5 HP
11 kWrated air-cooledair or water · 8 HP
13.5 kWrated air-cooledair or water · 8 HP
250L
10 kWrated air-cooledair or water · 5 HP
23 kWrated air-cooledair or water · 12 HP
25 kWrated water-cooledwater only · 12 HP
500L
13 kWrated air-cooledair or water · 8 HP
29 kWrated water-cooledair or water · 12 HP
44 kWrated water-cooledwater only · 15 HP
1000L
28.5 kWrated water-cooledair or water · 12 HP
47 kWrated water-cooledwater only · 20 HP
65.5 kWrated water-cooledwater only · 25 HP
Air or water cooled, your choiceWater cooled onlyEach power figure is rated for the cooling type under it; the other cooling type has a different rating, so ask us for it.

Air cooled or water cooled

The choice follows the ramp rate you need and the services in your building. Air cooling uses built-in fans and finned-tube condensers; water cooling moves heat from the refrigerant into an external water loop and suits larger chambers and faster rates.

Criterion
Air cooled
Water cooled
Initial infrastructure cost
None: plug in and run
Requires chiller
Ambient sensitivity
Highly dependent on room temperature
Independent of indoor room temperature
Indoor heat and noise
High thermal output and noise indoors
Low noise, heat rejected outdoors
Maintenance burden
Minimal (filter cleaning)
Moderate (water quality and scale management)
Heat dissipation
Cooling efficiency drops significantly if lab ambient rises above about 30 °C
Stays stable whatever the room temperature
Water supply
No water needed
Requires purified or deionized (DI) water
Fixed footprint
Easy to relocate within the facility
Constrained by water supply and return lines

Temperature, humidity and time on one controller

Set temperature, humidity and time together for every step. The chamber holds humidity without producing condensation inside, and the cascade refrigeration, SUS304 inner chamber and controller are built to run for weeks without a break.

Humidity management

20 to 98 %RH with no condensation produced in the chamber.

Temperature control

Anywhere from −70 °C to 150 °C, held within ±0.5 °C.

One controller

Set temperature, humidity and time from the 7-inch touch screen.

Temperature cycling

Average ramps of 5, 10 or 15 °C/min, set step by step.

Cooling system

Two-stage cascade compression cooling pulls the workspace down to −70 °C.

Integrated chamber

SUS304 mirror stainless inner chamber, insulated with ultra-fine glass wool and polyurethane foam.

Continuous testing

Runs for weeks without a break, with the test curve logged the whole time.

What stress screening catches

Fast temperature change finds the weak parts of a product long before a customer does.

HOT DWELLCOLD DWELLESSASTBurn-in

One chamber, many names

A rapid temperature change (humidity) chamber is named after the screening method it runs. The same chamber covers all of them.

  • ESSEnvironmental Stress Screening
  • ASTAccelerated Stress Testing
  • Burn-intest
Fast temperature change + condensation

Simulate harsh environments

Run fast temperature change (ESS) and condensation tests, with either the temperature test or the average temperature test method.

Screened outEarly failuresUseful lifeWear-outFailure rateProducts ship here

Accelerate product reliability

Weak parts fail early. Fast temperature swings bring those early failures forward into the lab, so the products that ship sit in the long, flat part of the curve. Solder fatigue life falls steeply as the temperature swing grows (the Coffin–Manson relation), which is why screening profiles use wide, fast swings. The chamber meets the stress-screening requirements for electronic equipment in these documents.

MIL-STD-2164MIL-HDBK-344ANAVMAT P-9492GJB 1032-90IPC-9701

Water recirculation: a closed loop

Condensate from the chamber drains back to the humidifier tank by gravity, so most of the water is used again. That keeps long, unattended high-humidity tests running and cuts DI / RO water use.

steamhumid airdrips backby gravityfeedWorkspacesteam humidifies the airHumidifierboils water to steamCooling coilcondenses water outWater tankRO or DI waterOverflow outlet: surplus only, to a traysteamhumid airdrips backby gravityfeedWorkspacesteam humidifies the airHumidifierboils water to steamCooling coilcondenses water outWater tankRO or DI waterOverflow outlet: surplus only, to a tray
No water line
the humidifier runs from a built-in tank
Water
RO or DI only
Reuse
condensate drains back into the tank by gravity
Long tests
less refilling during unattended high-humidity runs

Fill the tank with RO or DI water only; never use tap water.

Inside the heating, cooling and electrics

Heaters, refrigeration and electrics chosen for fast, repeatable cycling, in a cabinet insulated with ultra-fine glass wool and polyurethane foam.

Even heat, no hot or cold spots

Ceramic heating

Ceramic heating elements transfer heat quickly. PID control holds the fluctuation to ±0.5 °C and the deviation across the workspace to ±1.0 °C.

High stageheat to roomLow stageto −70 °CCascadeexchangerMechanical binary refrigeration

Refrigeration

Two compressor stages in a mechanical binary (cascade) system: the high stage cools the low stage, which reaches −70 °C and carries ramps of up to 15 °C/min.

SchneiderMitsubishiCarlo GavazziRainbowDanfossSAGINOMIYA

Electrical

Circuits laid out to USA standards with coded wiring and diagrams. Components from Schneider, Carlo Gavazzi, Mitsubishi and Rainbow; Danfoss and SAGINOMIYA solenoid valves.

Thirteen layers of protection

Separate protections cover the sample, the refrigeration system and the operator. On a fault the chamber shuts down and the circulation fan runs on for the delay you set. Remote control from a PC or phone, with remote assistance, lets you check a running test away from the lab.

  • Test sample over-temperature protection
  • Circulation motor overload protection
  • Compressor overload protection
  • Compressor over-pressure protection
  • Heater / humidifier dry-burn protection
  • Water shortage protection
  • Input power missing-phase and reverse-phase protection
  • Independent protection switches in the secondary circuit for lamp and anti-condensation wiring
  • Electric leakage protection
  • Three-color light alarm
  • Test sample terminal and equipment chain-control protection
  • Fault shutdown with circulation fan delay stop, settable 0–50 min (default 5 min)
  • Flame-retardant non-metallic materials: wire, cable, terminals, insulation
AlarmWarningRunning13protections

Programming and data

A 7-inch touch screen with time-segment and real-time programming, and a test curve record with no time limit.

Show all 15 features: Programming · Monitoring · Construction

Programming

  • Time-segment and real-time programming
  • 100 × 100 customizable programs
  • Separate variable rates per screening profile
  • Temperature, humidity and time set together
  • Temperature test and average temperature test methods

Monitoring

  • Complete real-time test curve analysis, no time limit
  • Troubleshooting system with alarms
  • Remote control via PC and phone
  • Remote assistance feature
  • Three-color light alarm

Construction

  • SUS304 mirror stainless steel inner body
  • Ultra-fine glass wool + polyurethane foam insulation
  • Special stainless-steel humidifier
  • Top-out, bottom-in forced air circulation
  • 2 shelves (100 / 250 L) or 3 shelves (500 / 1000 L)

Reference standards

Test methods and chamber specifications this chamber is designed to.

Show all 17 standards: IEC 60068, GB/T 2423, GJB 150
  • IEC60068-2-1.1990Test Methods for Low Temperature Test Chambers
  • IEC60068-2-2.1974Test Methods for High Temperature Test Chambers
  • IEC60068-2-1:2007 (GB/T2423.1-2008)Environmental testing – Part 2: Test methods – Tests A: Cold
  • IEC60068-2-2:2007 (GB/T2423.2-2008)Environmental testing for electric and electronic products Test methods – Tests B: Dry Heat
  • IEC 60068-2-14 2009Environmental testing – Part 2-14: Tests – Test N: Change of temperature
  • GB/T2423.22-2002Test Methods for Temperature Variations
  • GB/T2423.4-2008Test Methods for Cyclic Damp Heat Tests
  • GB/T2423.3-2006Test Methods for Thermal-humidity Test Chambers
  • GB/T10586-89Specifications for Thermal-humidity Test Chambers
  • GB/T2423.1-2008Test Methods for Low Temperature Test Chambers
  • GB/T2423.2-2008Test Methods for High Temperature Test Chambers
  • GB10592-89Specifications for Low/High Temperature Test Chambers
  • GB10589-89Specifications for Low Temperature Test Chambers
  • GB11158Specifications for High Temperature Test Chambers
  • GJB150.3High Temperature Tests
  • GJB150.4Low Temperature Tests
  • GJB150.9Damp Heat Tests

Questions buyers ask

Door access, options, warranty, returns, delivery and quotes.

Will it fit through my door?

Check the drawing above: each size is shown beside a standard 36 × 80 in door. The 5 °C/min water-cooled versions are the shallowest; crate sizes are in the specification table.

What does the warranty cover?

A 3-year worldwide warranty from the date of purchase, covering core components and controls. We repair first; if a major defect can’t be repaired, we replace the unit or give a full refund. Seals, heating tubes and other wear parts are excluded. A 2-year extended warranty can be added in the configurator.

Can I add shelves or a larger test hole?

Yes. Additional shelves and a custom test hole can be added in the configurator. If a custom hole could affect the door or wall sealing, we contact you before production.

Can I return it?

Yes, within 30 days of delivery if unused and in original condition. For non-quality returns the buyer pays return shipping, a 3% restocking fee and the card processing fee. Defective or damaged items are covered in full.

How is it delivered, and when does it ship?

Freight, duties and taxes are paid to your loading dock; moving the chamber inside is not included. Your confirmed ship date is included in the order acknowledgement.

Can I get a formal quote for a purchase order?

Yes. Select your size and options, then click Get a quote for a PDF quotation with DDP shipping included.

Choose size, rate, cooling and voltage

The price, delivery date and PDF quote update with your choices.

Configure yours