Lab Shaker

LORDERAN Stackable Incubated and Refrigerated Shaker, Max 3-Tier Laboratory Orbital Shaker, 0-300rpm, 4°C-60°C Ultra-Low Noise Single-Axis Drive

SKU
2012R
Availability
In stock
Questions

Ask about this instrument

LORDERAN Stackable Incubated and Refrigerated Shaker, Max 3-Tier Laboratory Orbital Shaker, 0-300rpm, 4°C-60°C Ultra-Low Noise Single-Axis Drive

An engineer replies by email or phone, usually within 12 hours.

Model · 2012R
Power · 220V

A stackable incubated and refrigerated orbital shaker for high-throughput life science research and bioprocess development. Stack two or three units in the floor space of one; each holds 4 to 60 °C, down to 18 °C below room temperature, with orbital or reciprocating shaking depending on the model.

37.0 °C200 rpm37.0 °C200 rpm37.0 °C200 rpm4–60 °CStack ×30–300 rpm
4–60 °Crefrigerated and heated
0–300 rpmspeed, ±1 rpm
±0.1 °Ctemperature accuracy
3 highstackable
84–180 Lfour models
Protein expressionEnzyme kineticsBacterial cultureThermophilesBioprocess development

How many units can I stack?

Two or three on the floor without loss of performance, for up to three times the capacity of a floor shaker in the same footprint.

How cold can it go?

Down to 4 °C, and up to 18 °C below room temperature. It heats to 60 °C.

Orbital or reciprocating?

The 2012R is orbital. The 2013R, 2019R and 2020R offer orbital and reciprocating shaking.

Which power does it need?

220 V or 110 V for every model. The 2013R and 2019R use an external transformer for 110 V.

Stackable Incubated and Refrigerated Shaker datasheetAll four models, plate sizes, temperature performance and power data on one sheet to share with your team.Download datasheet

Cold enough for proteins, warm enough for thermophiles

A CFC-free compressor system pulls the chamber down to 4 °C for low-temperature protein expression, sample preservation and enzyme kinetics. Heating takes it up to 60 °C for thermophilic bacteria, nucleic acid hybridization and vector construction. A 3D air duct keeps the temperature even.

0 °C60 °C

Cooling reaches 18 °C below room temperature, down to a minimum of 4 °C.

±0.1 °Ctemperature accuracy
±0.5 °Cuniformity at 37 °C
CFC-freecompressor system
2-panetempered glass viewport

Specifications

Models 2012R, 2013R, 2019R and 2020R. 220 V or 110 V; the 2013R and 2019R use an external transformer for 110 V.

Differs between models

Model 2012R 2013R 2019R 2020R
Model and shaking
Capacity 180 L 165 L 84 L 122 L
SKU 2012R 2013R 2019R 2020R
Shaking Mode Orbital Orbital & Reciprocating Orbital & Reciprocating Orbital & Reciprocating
Speed Range 0–300 rpm
Speed Accuracy ±1 rpm
Orbit / Stroke Dia. 25 mm
Programming Function Programing of repetition, steps and eight curves settings
Timing Range 0 – 999h59mins / continuous
Temperature
Temperature Range 4°C-60°C (RT-18°C, min 4°C)
Temperature Accuracy ±0.1°C (Constant temperature)
Temperature Uniformity ±0.5°C at 37°C
Size, weight and power
Size of shaking Plate 780×480 mm 480×440 mm 420×400 mm 542×445 mm
Inner Size 850×560×395 mm 560x520x565 mm 495×455×370 mm 610×505×400 mm
External Size 1150×800×640 mm 650×845×810 mm 580×740×580 mm 910×740×640 mm
Net Weight 320 kg 205 kg 150 kg 220 kg
Max Power 1720 W 1530 W 1030 W 1530 W
Rated current 7.8 A(220 V), 16 A(110 V) 7 A(220 V), 14 A(110 V) 4.7 A(220 V), 10 A(110 V) 7 A(220 V), 14 A(110 V)
Power Supply AC 220 V, 50/60 Hz; 110 V, 60 Hz AC 220 V, 50/60 Hz; 110 V, 60 Hz (110 V External transformer) AC 220 V, 50/60 Hz; 110 V, 60 Hz (110 V External transformer) AC 220 V, 50/60 Hz; 110 V, 60 Hz

A drive you will not hear

Brushless DC motors on heavy-duty bearings turn a central eccentric shaft, while four perimeter points carry the platform. The drive stays quiet even with uneven loads, needs no belts, and sits in a sealed housing away from heat and humidity.

Top view of the platform1Central eccentric shaftdrives the orbit4Perimeter support pointscarry uneven loads quietly0Beltsnothing to wear or slip1Sealed drive housingisolated from humidityØ 25 mm orbit
0–300 rpmspeed range
±1 rpmspeed accuracy
Ø 25 mmorbit
2 modesorbital or reciprocating*

Speed and orbit set the mixing

How hard the platform pushes your flasks depends on speed and orbit: the peak acceleration grows with the square of the speed and in proportion to the orbit. At 300 rpm a 25 mm orbit reaches about 1.26 g.

25 mm orbit0 g0.5 g1 g1.5 g0100200300speed, rpm1.26 g25 mm orbit0 g0.5 g1 g1.5 g0100200300speed, rpm1.26 g25 mm orbitup to 300 rpm: 1.26 g peak

Peak acceleration = (2π × rpm / 60)² × orbit radius, in multiples of g (9.81 m/s²). The drawing shows the orbits to scale.

Choose a model

Plate size and chamber volume set the load. Front elevations of a single unit, drawn to the same scale. *The 2013R, 2019R and 2020R switch between orbital and reciprocating.

2012R · 180L1150 mm780×480 plate2013R · 165L650 mm480×440 plate2019R · 84L580 mm420×400 plate2020R · 122L910 mm542×445 plate

Designed for long, unattended runs

Quiet, stable and easy to load, day after day.

Single-axis balanced drive

A balanced transmission with a variable-frequency drive: ultra-quiet and low-vibration.

Soft start

Controlled acceleration so flasks never splash.

Stackable to three

Stack two or three units on the floor without losing performance.

Drop-down door and pull-out tray

Load and unload vessels at a comfortable height.

7-inch colour touch screen

Program repetition, steps and eight curves, with USB data export.

Parameter lock

Encrypted parameters prevent accidental changes.

Auto-resume

Runs continue automatically after a power cut.

304 mirror stainless interior

Rounded corners for easy cleaning and contamination control.

Find the right LORDERAN shaker

Stack them, add CO₂, or choose a single floor or benchtop unit.

Shaker
Stackable refrigeratedThis product
Format
Stack up to 3 high
Stack up to 3 high
Floor-standing
Benchtop
Temperature
4–60 °C
4–60 °C
4–60 °C
4–60 °C
CO₂ and humidity
No
CO₂ 0.1–20 %, up to 95 %RH
No
No
Chamber
84–180 L
170–282 L
186 / 245 L
80 L

Questions buyers ask

Warranty, returns, delivery and quotes.

What does the warranty cover?

A 3-year worldwide warranty from the date of purchase, covering core components and controls. We repair first; if a major defect can’t be repaired, we replace the unit or give a full refund. Seals, belts and other wear parts are excluded.

Can I return it?

Yes, within 30 days of delivery if unused and in original condition. For non-quality returns the buyer pays return shipping, a 3% restocking fee and the card processing fee. Defective or damaged items are covered in full.

How is it delivered, and when does it ship?

Freight, duties and taxes are paid to your loading dock; moving the shaker inside is not included. Your confirmed ship date is included in the order acknowledgement.

Can I get a formal quote for a purchase order?

Yes. Select your size and options, then click Get a quote for a PDF quotation with DDP shipping included.

Choose model and voltage

The price and PDF quote update with your choices.

Configure yours