Biological Safety Cabinet

LORDERAN Class II A2 Biosafety Cabinet 2ft Benchtop, 99.9995% ULPA Filter, UV Lamp, Compact BSC for Lab & Hospital

SKU
700IIA2-Z-2
Availability
In stock
Questions

Ask about this instrument

LORDERAN Class II A2 Biosafety Cabinet 2ft Benchtop, 99.9995% ULPA Filter, UV Lamp, Compact BSC for Lab & Hospital

An engineer replies by email or phone, usually within 12 hours.

Power · 110V

A Class II A2 biological safety cabinet for work with infectious agents and cell culture. Inflow at the front at 0.53 m/s protects you, ULPA-filtered downflow at 0.33 m/s protects the sample, and the 30% that leaves passes a second ULPA filter, so no duct is needed. A compact 2 ft benchtop cabinet with a motorized front window. It is not for volatile toxic chemicals or radionuclides; for those, choose a B2.

LORDERAN0.53 m/sULPA 99.9995%Class II A20.53 m/s
0.53 m/sinflow velocity
0.33 m/sdownflow velocity
99.9995%ULPA at 0.12 μm
70 / 30% recirculated / exhausted
2 ftbenchtop width
Pharmaceutical developmentClinical researchMicrobiology labsCell culturePharmacy compounding

Do I need a duct?

No. 70% of the air is recirculated and 30% leaves through an ULPA exhaust filter, so the cabinet runs without a duct.

A2 or B2?

A2 for biological work. If the work also involves volatile toxic chemicals or radionuclides, choose a B2 total-exhaust cabinet.

When do the filters need changing?

Pressure sensors track filter life, and the controller warns you with an audible and visual alarm.

Which power does it need?

110 V or 220 V. Choose the voltage in the configurator.

Benchtop Biological Safety Cabinet CatalogEvery model, specification and option in one document to share with your team or purchasing office.Download catalog

Three protections in one airflow

Inflow at the front shields you. ULPA-filtered downflow keeps the sample clean. The 30% that leaves passes an exhaust ULPA filter, so the lab stays clean too, with no duct required.

ULPA supply filterRoom air in: inflow 0.53 m/sULPA-filtered downflow 0.33 m/s30% exhausted via ULPA · 70% recirculated
  • Inflow and downflow are monitored in real time by an airflow sensor.
  • Alarms sound for abnormal airflow, filter change and an unsafe window height.
  • Interlocks stop the UV lamp while the window is open or the LED lamp is on.
  • Filter life is tracked by pressure sensors.
ISO Class 3work zone with ULPA
200 mmtested safe opening
380 mmmaximum opening
≤ 65 dB(A)noise

Work at the tested opening

The protective air curtain is set and tested with the front window at 200 mm. Open it wider only to load or clean; the controller sounds an alarm when the window is above the safe height. The UV lamp runs only with the window closed.

ULPAUV on200 mm380 mmAlarmWork at 200 mm (8 in)Inflow is tested here: 0.53 m/sRaise to 380 mm to load or cleanAlarm sounds above the safe heightClose it for UVUV lamp is interlocked with the windowULPAUV on200 mm380 mmAlarmWork at 200 mm (8 in)Inflow is tested here: 0.53 m/sRaise to 380 mm to load or cleanAlarm sounds above the safe heightClose it for UVUV lamp is interlocked with the window

Where the airflow figure comes from

Air enters through the front opening at the inflow speed, and the same volume leaves through the exhaust ULPA filter. Opening area × speed gives about 230 m³/h for the 700lIA2-Z, the figure in the specifications.

interior width 600 mm200 mmfront windowtested opening 200 mm, air drawn in at 0.53 m/sOpening0.6 m × 0.2 m = 0.120 m²× inflow speed0.53 m/s = 0.0636 m³/s× 3600 s per hour≈ 229 m³/hListed for 700lIA2-Z230 m³/hinterior width 600 mmfront windowtested opening 200 mm, air drawn in at 0.53 m/sOpening0.6 m × 0.2 m = 0.120 m²× inflow speed0.53 m/s = 0.0636 m³/s× 3600 s per hour≈ 229 m³/hListed for 700lIA2-Z230 m³/h

The same calculation reproduces the listed airflow for every width.

Specifications

Model 700lIA2-Z, available in 110 V and 220 V versions.

Download the PDF
Model 700lIA2-Z
Dimensions
External Dims (W × D × H) 700 × 650 × 1280 mm
Internal Dims (W × D × H) 600 × 500 × 540 mm
Tested Opening Safety height: 200 mm (8 in)
Max Opening 380 mm
Net Weight 130 kg
Gross Weight 145 kg
Package Size (W × D × H) 760 × 787 × 1500 mm
Airflow and filtration
CFM 230 m³/h
Inflow Velocity 0.53 ± 0.025 m/s
Down Flow Velocity 0.33 ± 0.025 m/s
Airflow System 70% air recirculation, 30% air exhaust
ULPA Filter Two, 99.9995% efficient at 0.12 μm. Filter life indicator.
Cabinet and electrical
Front Window Motorized. Two-layer laminated toughened glass 5 mm. Anti UV.
Noise ≤ 65 dB(A)
Consumption 700 W
Socket Two, total load of two sockets: 500 W
Standard Accessory LED lamp*2, UV lamp*1, Remote control, Drain Valve*1, Waterproof Sockets*2, Foot Switch

Built for safe, comfortable daily work

The features you use every session, designed to keep the cabinet safe and simple to run.

Benchtop footprint

A full Class II A2 cabinet that sits on an ordinary lab bench and saves floor space.

ULPA filtration

Two ULPA filters at 99.9995% for 0.12 μm particles: about 10 times the efficiency of HEPA, for an ISO Class 3 work zone.

Real-time display

Filter life, fan run time and UV lamp run time, with an alarm below 10% remaining filter life.

Motorized front window

Raise and lower the sash with one hand, and clean the upper glass easily.

Time reserve

Schedule start-up and sterilizing so you skip about 30 minutes of waiting before and after work.

Audible and visual alarms

Warnings for abnormal airflow, filter change and an unsafe front window height.

Safety interlocks

UV lamp with front window, UV lamp with LED lamp, and blower with front window.

Remote control

Run the main functions from up to 6 meters away.

What's in the box

Delivered ready to work, with the accessories below as standard.

LED lamp*2UV lamp*1Remote controlDrain Valve*1Waterproof Sockets*2Foot Switch

Which biosafety cabinet is right for you?

Every LORDERAN Class II cabinet protects the operator, the sample and the environment. Choose by airflow type, size and controller.

Type
Class II A2This product
Class II B2
Air recirculated
70%
0%
Air exhausted
30%, through an ULPA filter
100%, through a dedicated duct
Exhaust connection
Not required (ductless)
Exhaust duct to outside
Typical work
Biological work without volatile or toxic chemicals or radionuclides
Biological work that also involves volatile toxic chemicals or radionuclides
Protects
Operator, sample and environment
Operator, sample and environment

Questions buyers ask

UV, warranty, returns, delivery and quotes.

What does the warranty cover?

A 3-year worldwide warranty from the date of purchase, covering core components and controls. We repair first; if a major defect can’t be repaired, we replace the unit or give a full refund. ULPA and HEPA filters, seals and other wear parts are excluded.

When can the UV lamp run?

Between sessions, to disinfect exposed surfaces. Interlocks stop it while the window is open or the LED lamp is on, so it runs only with the window closed and the work light off.

Can I return it?

Yes, within 30 days of delivery if unused and in original condition. For non-quality returns the buyer pays return shipping, a 3% restocking fee and the card processing fee. Defective or damaged items are covered in full.

How is it delivered, and when does it ship?

Freight, duties and taxes are paid to your loading dock; moving the cabinet inside is not included. Your confirmed ship date is included in the order acknowledgement.

Can I get a formal quote for a purchase order?

Yes. Select your size and options, then click Get a quote for a PDF quotation with DDP shipping included.

Download documents

Catalogs and manuals as PDF files.

Choose your voltage

The price and PDF quote update with your choices.

Configure yours