How to Use a Lab Centrifuge: Setup, Balancing and a Safe Run
Last updated November 2026 · LORDERAN how-to guide
Short answer: set the centrifuge on a solid, level bench with 300 mm (about 12 in) clear on every side. Seat the rotor on a clean taper and tighten the nut. Load tubes symmetrically, with opposite positions within 0.1 g of each other. Fill to no more than 90 % in swing-out rotors and 80 % in fixed-angle rotors. Then select the rotor, set × g, time and temperature, close the lid until it clicks and stay until the run is at speed and quiet. The figures in this guide come from the LORDERAN TDL-5A and TGL-21M manuals. Check the manual for your model; all ten are linked at the end.
1. Place it
A centrifuge pushes warm air out of its vents and needs room to shed it. The clear zone is also a safety zone: nobody should lean on the instrument or stand beside it longer than needed while it runs.
-
1Solid, level benchIt must carry the instrument, rotor and load without flexing, with every rubber foot touching: at least 80 kg (176 lb) plus load for the TDL-5A, 100 kg (220 lb) for the TGL-21M.
-
2300 mm (12 in) clear on every sideAway from pumps, shakers and other vibration, direct sun and heat sources.
-
3Its own grounded outletAt the voltage on the rating plate. No extension cords or power strips.
-
4Two hours to acclimateAfter cold transport or storage, wait at least 2 hours before switching on so condensation can dry.
2. Fit the rotor
A rotor that is not fully seated, or a loose nut, is a common cause of vibration. The fit takes a minute and is worth checking before every run.
-
1Clean the taper and the boreWipe both with a soft cloth and apply a very thin film of the supplied lubricant.
-
2Lower it straight downHold the rotor level and let the bore seat fully on the taper. Never drop or force it.
-
3Tighten the nut clockwiseHold the rotor still with the other hand. On fixed-angle rotors, fit and tighten the rotor lid too.
-
4Hang every bucketOn swing-out rotors all buckets go on, empty ones included, and each must swing freely.
To remove it, switch off first, hold the rotor, loosen the nut counterclockwise and lift it out vertically. Tilting it can bend the motor shaft.
3. Balance the load
The step that matters most. An uneven load makes the rotor shake, wears the drive and stops the run. Weigh the tubes; do not judge by eye. Opposite positions must match to within 0.1 g, with the same tube type and the same sample in each.
Balanced
Four equal tubes spaced 90° apart.
Unbalanced
Two tubes side by side, nothing opposite.
Loading patterns for a 12-place rotor
The tubes must be spread evenly around the axis. When the count does not divide evenly, add a blank tube of water at the same mass.
Fill limits
Fill tubes to no more than 90 % of their nominal volume in swing-out rotors and 80 % in fixed-angle rotors. Close every tube tightly and dry the outside.
The speed limit is the lowest of the rotor, bucket, adapter and tube limits. For commercial tubes, use the speed or × g the tube maker states. Samples denser than usual, such as concentrated salt or sucrose, need a lower speed: ask LORDERAN for the rotor's rated density.
4. Set up the run
-
1Select the rotorThe number on the screen must match the rotor fitted; the speed limits come from it.
-
2Enter × g or rpmSwitch Speed / RCF to enter × g directly. The instrument works out the rpm for that rotor. See RPM vs RCF.
-
3Set the timeBy default the timer starts when the rotor reaches set speed.
-
4Set the temperature−20 to +40 °C on refrigerated models.
-
5Set ACC and DECACC 1–10, DEC 0–10. DEC 0 lets the rotor coast to a stop.
-
6Close the lid and startPress both sides until it clicks. Stay until the rotor reaches set speed and runs quietly.
Example: serum gel tubes at 1300 × g for 10 minutes in the TDL-5A 18 × 7 ml rotor (r = 179 mm).
5. Finish safely
A beep marks the end of the run. The lid stays locked until the rotor has stopped.
-
1Wait for a full stopThen tap Open and lift the lid with both hands.
-
2Lift tubes out uprightSo the pellet is not disturbed. Check for spills or broken glass.
-
3Dry the chamberAfter cold runs, wipe it dry and leave the lid open for a while.
-
4End of the dayTake the rotor out, store it dry and switch off at the power switch.
When an alarm sounds
The screen turns red, the buzzer sounds and the drive stops. Tap the red area to acknowledge, fix the cause, then restart. Note the code and time; alarms are kept in Run Records.
Cleaning and rotor life
Clean with a soft cloth and a mild neutral detergent. Disinfect with an agent safe for aluminum and paint, such as 70 % ethanol. Do not use chlorine bleach: it corrodes aluminum. Wipe the touch screen with a dry cloth only.
A rotor that has been dropped, or shows cracks, corrosion pits, deformation or damaged hanger slots, stays out of use until LORDERAN has inspected it.
Manuals by model
Each manual has the rotor table, speed limits and alarm list for its model. Choosing a centrifuge? Start with RPM vs RCF: how to convert and choose or the centrifuge collection.
Questions
How exactly do the tubes need to be balanced?
The TDL-5A and TGL-21M manuals ask for opposite positions to match within 0.1 g. Weigh the filled tubes or buckets on a balance. Tubes that look equal often differ by more than that.
What if I have an odd number of tubes?
Add a blank tube filled with water to the same mass, so the count fits an even pattern. In a 12-place rotor, three tubes can also run alone if they sit 120° apart, in every fourth hole.
Do empty buckets have to be in the rotor?
Yes. On swing-out rotors every bucket and rack goes on for every run, empty ones included, and opposite buckets must be identical and swing freely.
Can I clean the rotor with bleach?
No. Chlorine-based disinfectants such as sodium hypochlorite corrode aluminum and paint. Use a mild neutral detergent, rinse and dry, and disinfect with 70 % ethanol.
How long does a rotor last?
Five years of service in these manuals. Then return it for inspection or replace it. Write the first-use date on the rotor case so you know when it is due.
Why won't the lid open?
It stays locked until the rotor has stopped. After a power failure, unplug the centrifuge, wait until the rotor is still and use the emergency release as your model's manual describes. Never release the lid while the rotor turns.