LORDERAN Horizontal Steam Sterilizer Autoclave, Gravity Displacement, Door Interlock, for Lab, Dental, Veterinary & Mushroom
- SKU
- 150J
- Availability
- In stock
- Questions
A horizontal steam sterilizer from 150 to 400 L for high-volume labs, clinics and mushroom growers. It uses gravity displacement: steam enters the chamber and pushes the heavier cold air out through the drain. Sterilize at any set point from 115 to 134 °C, then dry dressings and instruments in the same chamber. It needs a three-phase supply; interlocks keep the door shut while the chamber is under pressure.
Set the sterilization temperature anywhere from 115 to 134 °C. The vessel is designed for 142 °C and 0.28 MPa.
How is air removed?
By gravity displacement: steam enters from the top and pushes the heavier cold air out through the drain.
Which power does it need?
Three-phase 220 V or 380 V, 50/60 Hz; 12 to 18 kW by size.
Can it dry the load?
Yes. Drying runs in the same chamber after sterilizing, with a timer from 0 to 999 min.
When does the door open?
Only below 0.027 MPa. The cycle is blocked unless the door is closed and locked.
150J–400J Horizontal Steam Sterilizer datasheet (6 pages)
Full specifications, the sterilization cycle, saturated steam and F0 reference values, installation and warranty in one document to share with your team.Download datasheetOne cycle, four stages
Steam heats the load and drives cold air out of the chamber, the temperature holds at the set point for the sterilization time, pressure is released, and the drying stage removes moisture from dressings and instruments.
Temperature through a typical 134 °C cycle. Times are set by the user.
- Door stays locked until chamber pressure drops below 0.027 MPa.
- A cycle cannot start unless the door is securely closed.
- Water shortage protection cuts power and sounds an alarm.
- Automatic over-temperature and over-pressure protection.
Why it runs under pressure
Water boils at 100 °C at normal air pressure. To reach 121 °C or 134 °C, steam has to be held above atmospheric pressure, which is why the chamber is a pressure vessel. Air has to be removed first: an air and steam mixture at the same pressure is cooler than pure steam. At 134 °C the steam needs about 0.20 MPa, below the 0.22 MPa this unit is rated for.
Saturated steam pressure calculated with the IAPWS-IF97 equations, shown above atmospheric pressure (gauge). Pure steam with all air removed.
Choose by chamber size
All four models share the same pressure rating, temperature range and controls. Chamber diameter grows from 472 to 620 mm.
| Model | 150J | 200J | 300J | 400J |
|---|---|---|---|---|
| Volume | 150 L | 200 L | 300 L | 400 L |
| Cavity (mm) | φ472x850 | φ472x1130 | φ620x960 | φ620x1280 |
| Basket (mm) | 780x452x217 | 1060x452x217 | 860x600x285 | 850x678x330 |
| Overall (mm) | 1320x635x1520 | 1600x635x1520 | 1540x790x1700 | 1570x875x1780 |
| Power | 12 kW | 15 kW | 18 kW | 18 kW |
How gravity displacement works
No vacuum pump is needed. Steam does the work, because it is lighter than air.
Fill
Water from the supply at 0.1–0.2 MPa is heated to make steam inside the vessel.
Purge
Steam fills the chamber from the top and pushes cold air down and out of the drain.
Sterilize
The load holds at 115–134 °C for the time you set, up to 999 minutes.
Exhaust and dry
Pressure is released, then the drying stage runs for up to 999 minutes.
Safety built into every step
A manual structure that operators know, with modern interlocks and protection.
Safety
- Door locked until pressure is below 0.027 MPa
- Cycle blocked unless the door is closed
- Water shortage protection with alarm
- Over-temperature protection
- Over-pressure protection
Process
- Gravity displacement air removal
- Sterilization temperature 115–134 °C
- Sterilization time 0–999 min
- Drying time 0–999 min
- Temperature uniformity ≤±2.0 °C
Build
- High-quality stainless steel chamber
- Strong corrosion resistance
- Horizontal loading with baskets
- Three-phase power supply
- Drying for dressings and instruments
Specifications
Models 150J, 200J, 300J and 400J. "/" means not applicable to this gravity displacement design.
Differs between models
| Model | 150J | 200J | 300J | 400J |
|---|---|---|---|---|
| Main technical parameters | ||||
| Design pressure | 0.28MPa | |||
| Design temperature | 142°C | |||
| Working pressure | 0.22MPa | |||
| Operating temperature | 115-134°C | |||
| Temperature uniformity | ≤ ±2.0°C | |||
| Water source pressure | 0.1 to 0.2MPa | |||
| Vacuum | / | |||
| Compressed air pressure | / | |||
| External steam pressure | / | |||
| Power supply voltage | (220 V or 380 V), Three-phase 50/60 Hz | |||
| Product parameters | ||||
| Volume | 150 L | 200 L | 300 L | 400 L |
| Power | 12 kW | 15 kW | 18 kW | 18 kW |
| Voltage | AC 380 V/220 V 50/60 Hz (3-phases) | |||
| Rated working pressure | 0.22MPa | |||
| Rated working temperature | 134°C | |||
| Sterilization temperature selection range | 115 to 134°C | |||
| Sterilization time selection range | 0 to 999 min | |||
| Drying time selection range | 0 to 999 min | |||
| Dimensions and weight | ||||
| Cavity dimensions (mm) | φ472x850 | φ472x1130 | φ620x960 | φ620x1280 |
| Basket dimensions (mm) | 780x452x217 | 1060x452x217 | 860x600x285 | 850x678x330 |
| Overall dimensions (mm) | 1320x635x1520 | 1600x635x1520 | 1540x790x1700 | 1570x875x1780 |
| Packaging dimensions (L×W×H) (mm) | 1470x785x1720 | 1750x785x1720 | 1690x940x1900 | 1830x1025x1980 |
| Gross/net weight | 350/270 kg | 410/310 kg | 600/450 kg | 680/510 kg |
Questions buyers ask
Warranty, returns, delivery and quotes.
What does the warranty cover?
A 1-year warranty from the date of purchase, covering core components and controls. We repair first; if a major defect can’t be repaired, we replace the unit or give a full refund. Seals, heating tubes and other wear parts are excluded.
Can I return it?
Yes, within 30 days of delivery if unused and in original condition. For non-quality returns the buyer pays return shipping, a 3% restocking fee and the card processing fee. Defective or damaged items are covered in full.
How is it delivered, and when does it ship?
Freight, duties and taxes are paid to your loading dock; moving the autoclave inside is not included. Your confirmed ship date is included in the order acknowledgement.
Can I get a formal quote for a purchase order?
Yes. Select your size and options, then click Get a quote for a PDF quotation with DDP shipping included.
Choose your size