How to Use a Biosafety Cabinet: Setup, Safe Work and Cleaning

November 21, 2026

Last updated November 2026 · LORDERAN how-to guide

Short answer: put the cabinet where doors, air vents and passing people will not disturb its air curtain. Before work, switch the UV lamp off, set the sash to its working height and let the blower run at least five minutes. Load everything first, work at least 4 inches in from the front grille, move your arms slowly and work from clean to dirty. Afterwards wipe the interior with 70 % ethanol or diluted bleach, and have the cabinet certified at least once a year. This guide follows the CDC/NIH BMBL, Appendix A, with figures from the LORDERAN Class II A2 Pro manual.

1. Place it

A Class II cabinet protects you with a curtain of room air drawn in across the front opening. Drafts from doors, supply vents, fans and people walking past can break that curtain.

  1. 1
    Away from the door and trafficThe ideal spot is remote from the entry, out of the walkway. At least 1.5 m (5 ft) from a doorway to the front opening.
  2. 2
    Away from moving airAt least 1.5 m (5 ft) from a supply air diffuser, and no open windows or portable fans nearby.
  3. 3
    Room in front and aboveAt least 2 m (6.5 ft) to an opposite wall, 3 m (10 ft) to an opposite cabinet, fume hood or exhaust outlet, and 12–14 in clear above the cabinet.
  4. 4
    Room conditionsIndoors, 15 to 35 °C (59 to 95 °F), no more than 75 %RH, on a grounded supply.
  5. 5
    Certify before useThe cabinet must be tested at installation, before it goes into service.
BSCfront openingDoor≥ 1.5 m (5 ft)≥ 2 mto a wallSupply diffuser≥ 1.5 m (5 ft)Keep traffic off the faceAlso 3 m (10 ft) to an opposite cabinet, fume hood or exhaustoutlet, and 12–14 in clear above the cabinet.

2. Protect the air curtain

Room air flows in through the front grille while filtered air falls over the work surface. Fast sweeping arm movements, or anything blocking the front grille, let air from inside the cabinet escape toward you.

Slow and straight

Arms move in and out perpendicular to the opening.

ULPA filtersashroom air inAir curtain holds

Sweeping, grille blocked

A towel on the front grille and a fast sideways sweep.

ULPA filtersashroom air inAir curtain breaks
Room air inFiltered downflowAir escaping

3. Lay out the work surface

  1. 1
    Bring it all in firstEvery item you need goes in before you start, so your arms cross the opening as little as possible. Place materials as far back as practical.
  2. 2
    Clean to dirtyClean materials on one side, the work in the middle, waste on the other. Keep clean materials at least 1 ft (30 cm) from anything that makes aerosols.
  3. 3
    4 inches inDo all work at least 4 in (10 cm) behind the front grille.
  4. 4
    Grilles clearPlastic-backed absorbent toweling may cover the work surface, never the front or rear grille.
  5. 5
    Mind the loadThe A2 Pro manual allows up to 23 kg on any 25 × 25 cm area of the work surface.
rear grille: keep clearfront grille: keep clear4 in (10 cm) lineCleanWorkWasteWork moves one way, clean to dirty

4. A work session, step by step

BMBL asks for at least five minutes of blower time before work to purge the cabinet; the A2 Pro manual allows 30 minutes after switch-on. Use the longer time if your lab's procedure does.

Before

  1. 1
    UV off, PPE onUnshielded UV lamps must be off whenever people are in the room.
  2. 2
    Sash to 200 mm (8 in)Switch on the blower and light. The blower will not run with the sash fully closed.
  3. 3
    PurgeLet the blower run before you start and check the airflow display shows no alarm.
  4. 4
    Wipe and loadWipe the work surface with 70 % ethanol, then load everything you need.

During

  1. 1
    Wait a minuteAfter placing your arms inside, wait about a minute before you start working.
  2. 2
    Move slowlyIn and out perpendicular to the opening, never sweeping sideways.
  3. 3
    No open flamesThey are not needed in the cabinet and disturb the airflow. Use an electric (infrared) sterilizer.
  4. 4
    Sash stays putKeep it at the working height; the alarm sounds if it rises.

After

  1. 1
    Close everythingClose all containers before they leave the cabinet.
  2. 2
    Wipe down70 % ethanol, or 1:100 household bleach followed by sterile water to remove the chlorine.
  3. 3
    Blower on or offAfter the final wipe the blower may be left on or switched off. Lower the sash fully when you finish.
  4. 4
    UV, if usedNever as the only disinfection. On the A2 Pro it runs only with the sash closed and the blower and light off.

5. If something spills

Small spills inside a running cabinet can be dealt with straight away. Keep the blower running so the spill stays contained.

  1. 1
    Cover and disinfectCover the spill with absorbent material and apply disinfectant.
  2. 2
    Give it timeLeave it for the disinfectant's contact time, usually 20 to 30 minutes, then wipe up and wipe the surfaces again.
  3. 3
    Through the grilles?A spill large enough to run through the front or rear grille needs the area under the work surface decontaminated as well. Follow your lab's spill procedure.

Alarms

The A2 Pro gives an audible and visual alarm for each of these. Working velocities: inflow 0.53 m/s (104 fpm), downflow 0.33 m/s (65 fpm).

ALARMCAUSEWHAT TO DO
Sash heightThe front window is above the working height. The display flashes and the buzzer sounds.Bring the sash back to 200 mm (8 in).
AirflowInflow below 0.42 m/s (83 fpm) or downflow below 0.26 m/s (51 fpm), 20 % under the set values.Stop work, close up your materials and check the grilles are clear. If it persists, call service.
Filter pressureSupply or exhaust filter pressure outside the expected range.Stop work. The filter needs service; a cabinet used with infectious material is decontaminated before a filter change.

Cleaning and certification

Use disinfectants that will not damage 304 stainless steel, and non-abrasive cleaners on the glass and outside. Disconnect power before any maintenance.

AFTER EACH USEWipe it down70 % ethanol or 1:100 bleach on the work surface and walls; follow bleach with sterile water.
END OF THE DAYFinal wipeWipe all interior surfaces. The blower may then stay on or be switched off.
MONTHLYDeep cleanWork surface, interior walls and glass, leaving out the air grilles.
UV LAMPIf you use oneCheck its output; replace when it reads below 40 µW/cm² (BMBL) or after 600 h of use (A2 Pro manual).
YEARLYCertificationField test to NSF/ANSI 49 Annex F by qualified personnel, at least annually.
ALSORecertify afterInstallation, repair, relocation or a HEPA/ULPA filter change.

Manuals and guides

Each LORDERAN biosafety cabinet has its own manual with its velocities, alarms and filter details.

Every modelBiosafety cabinet manuals
Buying guideClass II A2 vs B2
Buying guideClean bench vs BSC
All modelsCollection

Questions

How long should the blower run before I start?

BMBL says at least five minutes, to purge the cabinet. The LORDERAN A2 Pro manual allows 30 minutes after switch-on. Follow whichever your lab's procedure sets.

Should I use the UV lamp?

Not as the only disinfection: BMBL says UV should not be the sole method. If you use it, keep it off while the room is occupied and replace the bulb when its output falls below 40 µW/cm².

Can I use a Bunsen burner inside?

No. BMBL says open flames are neither required nor recommended in a cabinet; they disturb the airflow. An electric (infrared) loop sterilizer does the job.

How often does the cabinet need certifying?

At installation, at least once a year, and again after repair, relocation or a HEPA or ULPA filter change. The field test follows NSF/ANSI 49 Annex F and is done by qualified personnel.

Ethanol or bleach?

Both appear in BMBL: 70 % ethanol, or a 1:100 dilution of household bleach. After bleach, wipe again with sterile water to remove the residual chlorine.

Why does the sash alarm keep sounding?

The front window is above its working height of 200 mm (8 in). Bring it back down; working with the sash higher weakens the inflow that protects you.