LORDERAN Temperature Humidity Test Chamber 0-100°C, 20-98% RH, Water Recirculation, for Electronics & Material Testing, HWS Series
- SKU
- HWS150
- Availability
- In stock
- Questions
Long, steady temperature and humidity tests: stability studies, shelf-life testing, material conditioning and bioproduct storage, at 0 to 100 °C and 20 to 98 %RH, in four sizes from 150 to 1000 L. Humidity is controlled from 15 to 85 °C, and the lowest humidity you can reach rises as the temperature falls, so check the control range chart for your conditions.
Will it run ICH conditions?
Yes. 25 °C / 60 %RH, 30 °C / 65 %RH and 40 °C / 75 %RH all sit inside the controlled zone, held for months.
How often do I add water?
Rarely. Condensate and unused water are pumped back to the humidifier, and a refill bottle tops up the loop.
Which voltage should I order?
Each size comes in 110 V and 208–240 V versions. At 220 V: 6.8 A for 150 and 250 L, 11.4 A for 500 and 1000 L.
Can it cool below room temperature?
Yes, down to 0 °C with humidity off. Humidity is controlled from 15 to 85 °C; see the chart.
How it holds one condition for months
Air circulates through a duct beside the workspace; water circulates through a loop below it.
Simplified section, not to scale; part positions are schematic. Screen values are an example.
-
1Load the samples.Two racks as standard. A 100 mm test hole takes cables to powered samples.
-
2Set the condition.Temperature and humidity on the touch screen; a PT100 platinum sensor reads the temperature.
-
3Condition the air.The fan drives air through the duct: the evaporator cools and dries it, the heater warms it, the humidifier adds moisture. It leaves evenly across the shelves.
-
4Reuse the water.Condensate and unevaporated water are collected and pumped back to the humidifier, and a refill bottle tops up the loop automatically. If water runs low, the humidifier heater switches off.
The stable damp-heat working zone
Temperature runs from 0 to 100 °C. Where temperature and humidity are controlled together, the chamber holds the coloured zone steadily for weeks. The three markers are the ICH stability conditions used in pharmaceutical stability programs; all three sit well inside the zone.
Specifications
Models HWS150, HWS250, HWS500 and HWS1000, in 110 V and 208–240 V versions.
Differs between models
| Model | 150 L | 250 L | 500 L | 1000 L |
|---|---|---|---|---|
| Model and size | ||||
| SKU | HWS150 | HWS250 | HWS500 | HWS1000 |
| Inner size (W × D × H) | 600 × 405 × 620 mm | 600 × 500 × 830 mm | 800 × 700 × 900 mm | 1000 × 1000 × 1000 mm |
| External size (W × D × H) | 830 × 800 × 1500 mm | 830 × 900 × 1700 mm | 1030 × 960 × 1790 mm | 1230 × 1400 × 1850 mm |
| Test Hole | 100 mm*1 | |||
| Racks | 2 | |||
| Weight | 170 kg | 190 kg | 320 kg | 400 kg |
| Temperature and humidity | ||||
| Temperature range | 0 to 100 °C | |||
| Humidity range | 20 to 98 %RH | |||
| Temperature fluctuation | ±0.5 °C | |||
| Humidity fluctuation | ±2 %RH | |||
| Temperature deviation | ±1.5 °C | |||
| Humidity deviation | ±3.0 %RH | |||
| Refrigeration, control and power | ||||
| Refrigeration method | High-efficiency direct cooling | |||
| Compressor | Hermetic compressor | |||
| Cooling method | Air cooled | |||
| Refrigerant | R290 | |||
| Temperature sensor | Platinum resistance PT100 | |||
| Display | Touch Screen | |||
| Power | 1.5 kW | 1.5 kW | 2.5 kW | 2.5 kW |
Choose a size
Same range, controller and protection on every size. Front elevations drawn to the same scale.
Ready for long-term stability studies
Typical ICH conditions for drug substances and products, all inside the controlled zone, held for months at low running cost.
Real-time stability at standard storage conditions.
When accelerated results show a significant change.
Faster degradation for early shelf-life estimates.
Testing frequency for a product with a proposed shelf life of at least 12 months: every 3 months in year 1, every 6 months in year 2, then yearly (36 months shown). Intermediate: 4 time points over 12 months; accelerated: 3 over 6 months. The long-term condition alone keeps a chamber busy for years.
Conditions as commonly specified in ICH Q1A(R2). Your protocol defines the exact conditions and durations.
Water recirculation: weeks without refilling
Condensate and unevaporated water are collected and pumped back to the humidifier reservoir, for more than 80% better water utilization, less DI or RO water, no external drain line and unattended running for weeks or months.
Engineered to run cheaply for months
What keeps running costs low, and the details that make long tests easier.
Variable-frequency compressor
Output follows the load instead of cycling at full speed, for steadier control and lower energy use.
Thicker insulation
Low-conductivity insulation reduces the work the heater and compressor must do.
Automatic water refill
A recirculation tank feeds the humidifier automatically.
Optimized air duct
Even airflow for better temperature and humidity uniformity.
SUS 304B interior
Acid- and corrosion-resistant stainless steel that wipes clean.
R290 refrigerant
A natural refrigerant, in an air-cooled hermetic system.
Touch screen, PT100
Clear touch control with a platinum resistance PT100 sensor.
100 mm test hole
Run cables to powered samples during the test.
Low current draw
Rated current by size and supply voltage, to plan the circuit with your facilities team. For a load that runs for days, keep it to 80% of the breaker rating.
Safety protection
The chamber protects itself, your samples and your lab.
Show all 4 safety protections
- Compressor overload protection
- Humidification heater dry-burn protection: power is cut when reservoir water is too low
- Over-temperature protection: heating, and if needed cooling, shut down
- Electrical short-circuit protection
Find the right LORDERAN environmental chamber
Steady climate, damp heat or fast ramps: choose by how your product must be tested.
Questions buyers ask
Humidity limits, warranty, returns, delivery and quotes.
Which humidity is possible at which temperature?
See the control range chart: humidity is controlled from 15 to 85 °C, up to 98 %RH between 15 and 60 °C and 90 %RH at 85 °C; the minimum humidity rises at lower temperatures. Outside the zone, ask our engineers.
What does the warranty cover?
A 2-year warranty from the date of purchase, covering core components and controls. We repair first; if a major defect can’t be repaired, we replace the unit or give a full refund. Seals, heating tubes and other wear parts are excluded.
Can I return it?
Yes, within 30 days of delivery if unused and in original condition. For non-quality returns the buyer pays return shipping, a 3% restocking fee and the card processing fee. Defective or damaged items are covered in full.
How is it delivered, and when does it ship?
Freight, duties and taxes are paid to your loading dock; moving the chamber inside is not included. Your confirmed ship date is included in the order acknowledgement.
Can I get a formal quote for a purchase order?
Yes. Select your size and options, then click Get a quote for a PDF quotation with DDP shipping included.