150L Vertical Sterilizer Autoclave

Two-year warranty. Digital display of working status. Automatic exhaust. Over temperature& overpressure auto-protection. Safe protection of water run out. Automatically shut off with beep reminding after sterilization. Comes With 2 stainless steel sterilizing baskets.
SKU: LS150B
Availability: In Stock
$4,074.00 AUD
$5,378.00 AUD
$4,074.00 AUD
Power: 220V/28A
Subtotal: $4,074.00

Free Shipping

Free shipping from China on all orders (Door to Door)

30 Day Money Back Guarantee

If approved, you’ll be automatically refunded on your original payment method.

What DDP Term means

Lorderan paying for all charges including freight, duties, taxes, customs clearance charges up to the customer's door, warehouse or fulfillment center.
150L Vertical Sterilizer Autoclave

150L Vertical Sterilizer Autoclave

$5,378.00 $4,074.00

150L Vertical Sterilizer Autoclave

$5,378.00 $4,074.00
Power: 220V/28A

    Compact, Powerful, Reliable

    Automatic high-pressure steam sterilization

    Full stainless steel 304 structure

    Plug & Play Installation

    Reliable Performance

    Excellent value for money

    Easy To Replace Parts

    DESIGN

    OPERATE

    SAFETY

    01

    Small foot print

    LED display

    Leakage breaker

    02

    Bolt structure

    up to 134℃

    Safety valve

    03

    Auto-exhaust 

    Easy setting

    Built-in voltage regulator

    04

    Full SS304

    Fully automatic

    Anti-dry burning

    05

    Large load

    Built-in cooling fan

    Overpressure protection

    The Critical Equation: Why 150L Autoclave Are Not Supplied for 110/120V Power

    Circuit and Breaker Limitations

    120V Constraints: Standard residential and light laboratory circuits in North America are typically protected by 15~20A circuit breakers.


    Electrical codes (like the US NEC) often require derating the breaker capacity to 80% for continuous running loads (over 3 hours). This means 50A would require a circuit breaker near 70A, which is highly unusual and costly for a single-phase 110V installation.

    Wiring and Material Cost

    Higher current requires significantly thicker gauge wiring to safely carry the 50A without overheating (typically 4 or 6 AWG wire, depending on factors).

    Single-Phase 208V/220V: This is the common laboratory/commercial standard, requiring a much more manageable circuit of around 30A.

    Validated Performance and Robust Quality Assurance

    Exhaust Mode

    Exhaust mode (automatic/non-vent) is instantly selectable via a one-button control.
    Provides a sealed environment for applications such as sterilizing liquids (preventing boil-over/rapid depressurization) or for specific research protocols requiring controlled, slow exhaust.

    Comprehensive Protection

    Over-temperature protection, over-pressure self-venting protection, low water level protection, dry-run protection.

    Control System

    Simply set the sterilization temperature and duration for fully automatic operation.

    Remove complexity and simplify, no threshold for use, dramatically reducing the failure rate.

    Heating Elements

    Independent audible and visual alarm system for each of the two heating elements for easy troubleshooting and replacement.

    Durability

    Under strict quality control to meet or exceed industry standards (e.g., EN 285 or ASME standards), assuring its quality and efficacy. The robust design and continuous monitoring systems ensure every cycle achieves a verified Sterility Assurance Level.

    Warranty

    Two-year warranty to make our users worry-free.

    View, Download user manual

    This guide will help you get started with Lorderan 150L autoclave.