Sample Preparation
Biological Safety Cabinet
Laminar Flow Cabinet
Fume Hood
Freeze Dryer
Steam Sterilizer
Vertical Pressure Steam Sterilizers
Lab Shaker
Open Air Shaker
Stackable Shaker
Floor Model Shaker
Test Chamber
Tablet Testing
Warranty
Support
FAQ
Blog
My Wish List




Free shipping on all orders (Door to Door)
We can ship to worldwide. Note that there are restrictions on some products, and some products cannot be shipped to international destinations.
Learn more from our shippling policy https://www.lorderan.com/policies/shipping-policy
Learn more from our warranty policy: https://www.lorderan.com/pages/warranty-policy
You may return most new, unopened items within 30 days of delivery for a full refund. We'll also pay the return shipping costs if the return is a result of our error (you received an incorrect or defective item, etc.).
You should expect to receive your refund within four weeks of giving your package to the return shipper, however, in many cases you will receive a refund more quickly. This time period includes the transit time for us to receive your return from the shipper, the time it takes us to process your return once we receive it, and the time it takes your bank to process our refund request (3 to 5 business days).
If you need to return an item, simply login to your account, view the order using the "Complete Orders" link under the My Account menu and click the Return Item(s) button. We'll notify you via e-mail of your refund once we've received and processed the returned item. Learn more from our refund policy https://www.lorderan.com/policies/refund-policy
Delivered duty paid (DDP Term) shipping is a type of delivery where Lorderan paying for all charges including freight, duties, taxes, customs clearance charges up to the customer's door, warehouse or fulfillment center.
Cost-effectiveness
Versatility
FourIN ONE
We can sterilize a wide range of materials and liquids without needing multiple devices.
Efficiency
Multi-FUNCTIONAL
We can seamlessly switch between modes depending on their needs, streamlining their workflow.
Convenience
EasilyCONTROL
A single, integrated interface makes it easier to operate and maintain the sterilizer.
Eco-Friendly & Mess-Free Operation
No External Steam Emission
Key safety and environmental feature.
Steam is not vented directly into the lab atmosphere during the pre-cycle air removal or the post-cycle depressurization.
Standard Drying Function
Significantly reduce or completely eliminate residual moisture on sterilized items to prevent secondary contamination or rusting after removal.
Built-in Steam Tank
All exhausted steam and air are channeled into this internal tank, where the steam is condensed back into water.
Maintaining a high level of safety and convenience for the operator and the laboratory environment.
Selecting The Optimal Sterilization Mode
This mode is designed for sterilizing solid, non-porous items such as metal instruments, glassware, and unwrapped tools. The cycle typically involves a rapid rise in temperature and pressure, followed by a sustained high-temperature phase to ensure thorough sterilization. This mode is ideal for items that do not readily absorb moisture.
This mode is specifically for sterilizing liquids such as culture media, solutions, and other fluids. It is crucial in this mode to prevent boil-over and ensure even heat distribution. The autoclave will often have a slower exhaust phase to avoid rapid pressure release, which could cause liquids to boil violently. This mode is essential for preparing sterile liquids for laboratory and medical use.
This mode is used for melting substances such as agar and then holding them at a specific temperature to prevent solidification. It is particularly useful in microbiology and other fields where maintaining liquids in a molten state is necessary for subsequent procedures. This mode allows for precise temperature control and extended holding times.
Customization Mode
This mode allows users to set their own sterilization parameters, including temperature, time, and pressure.
Multi-Functional Laboratory Autoclave
Versatile Multi-Mode Operation
(Solid, Liquid & Waste): Custom pressure-relief algorithms prevent liquid boil-over/spill while maximizing exhaust speed for solid tools and bags.
Automatic Keep-Warm Function
Seamlessly switches to a temperature-holding phase 45℃ to 60℃ after sterilization, keeping agar media liquid and ready for immediate plate pouring.
Dedicated Agar Melting Mode
Gently remelts solidified agar without subject it to repeated high-pressure cycles, preserving nutrients and preventing thermal degradation.
Unattended Overnight Operation
Enables hands-free sterilization cycles—run the machine at night and retrieve perfectly warm, liquid media the next morning.
1
Multi-stage exhaust mode
Once sterilization is complete, the exhaust valve can vent at the preset multi-stage exhaust rate, thereby shortening the duration of the entire sterilization process.
2
Memory storage system
Capable of memorizing and storing the operating parameters of the sterilization cycle.
3
Safety interlock device
The door remains securely locked as long as residual pressure or elevated temperature exists, completely preventing accidental opening, steam burns, and operator hazards.
1
2
3
4
5
6
4
Inspection interface
Equipped with temperature and pressure calibration ports for effortless field calibration, compatible with 3Q (IQ/OQ/PQ) validation adapters.
5
Steam internal circulation
Downward exhaust structure, automatically exhausts cold air during sterilization and hot air after sterilization is completed.
6
Built-in printer
Prints out real-time sterilization data—including date, time, cycle number, sterilization temperature, pressure, and hold duration.
| Sterilize temperature:106~134℃ | Sterilize time:1-999mins | Temperature resolution:0.1℃ |
| Melting temperature:60~100℃ | Holding temperature: 45~60℃ | Temperature fluctuation:≤0.5℃ |
| Melting time:1~999min | Holding time: 1~999min | Drying time:0~240min |
| Exhaust interval:0-250S | Exhaust time:0-60S | Exhaust times:0-9 |
| Displace temperature:100-110℃ | Design pressure:0.3MPa | Max working pressure:0.24MPa |
N-35X
N-50X
N-75X
N-100X
φ300x450mm
φ350x515mm
φ400x585mm
φ450x615mm
33psi (0.23mpa)
33psi (0.23mpa)
33psi (0.23mpa)
33psi (0.23mpa)
134℃
134℃
134℃
134℃
116~134℃ (240.8~273°F)
116~134℃ (240.8~273°F)
116~134℃ (240.8~273°F)
116~134℃ (240.8~273°F)
≤0.5℃
≤0.5℃
≤0.5℃
≤0.5℃
1~999mins
1~999mins
1~999mins
1~999mins
0~240mins
0~240mins
0~240mins
0~240mins
1*(φ275x395mm)
2*(φ325x235)
2*(φ375x265)
2*(φ425x275)
Automatic internal exhaust
Automatic internal exhaust
Automatic internal exhaust
Automatic internal exhaust
Manual drainage
Manual drainage
Manual drainage
Manual drainage
2.5kw, 12amps(220V), 21amps (120V)
3kw, 13.6amps(220V), 25amps (120V)
4.5kw, 20.5amps(220V), 37.5amps (120V)
4.5kw, 20.5amps(220V), 37.5amps (120V)
430x610x1030mm
500x675x1100mm
540x716x1165mm
575x748x1255
90/74kg
125/97kg
145/114kg
161/129kg
N-35X
φ300x450mm
33psi (0.23mpa)
134℃
116~134℃ (240.8~273°F)
≤0.5℃
1~999mins
0~240mins
1*(φ275x395mm)
Automatic internal exhaust
Manual drainage
2.5kw, 12amps(220V), 21amps (120V)
430x610x1030mm
90/74kg
N-50X
φ350x515mm
33psi (0.23mpa)
134℃
116~134℃ (240.8~273°F)
≤0.5℃
1~999mins
0~240mins
2*(φ325x235)
Automatic internal exhaust
Manual drainage
3kw, 13.6amps(220V), 25amps (120V)
500x675x1100mm
125/97kg
N-75X
φ400x585mm
33psi (0.23mpa)
134℃
116~134℃ (240.8~273°F)
≤0.5℃
1~999mins
0~240mins
2*(φ375x265)
Automatic internal exhaust
Manual drainage
4.5kw, 20.5amps(220V), 37.5amps (120V)
540x716x1165mm
145/114kg
N-100X
φ450x615mm
33psi (0.23mpa)
134℃
116~134℃ (240.8~273°F)
≤0.5℃
1~999mins
0~240mins
2*(φ425x275)
Automatic internal exhaust
Manual drainage
4.5kw, 20.5amps(220V), 37.5amps (120V)
575x748x1255
161/129kg
Thanks for subscribing!
This email has been registered!