How to Choose a Lab Autoclave: Size, Type and Cycles

November 18, 2026

Last updated November 2026 · LORDERAN buying guide

Short answer: size the autoclave by the largest load you run in one cycle, with room left for steam to reach every item. Most labs that sterilize media, glassware and waste choose a 50 to 100 L vertical autoclave. Choose a horizontal 150 L or larger for high volume, and plan for three-phase power and a water supply. Then pick features by load: liquids need a slow exhaust, instruments and glassware need drying, and sealed food packs need counter-pressure.

How steam replaces air

An autoclave sterilizes with saturated steam under pressure, usually at 121 °C (250 °F) or 132 to 134 °C. Air has to leave the chamber first. CDC notes that air that is not removed will interfere with steam contact, and an air and steam mixture at the same pressure is cooler than pure steam.

Vertical autoclave

Water in the base boils into steam. Cold air is purged through the exhaust before the hold starts.

Cold air out Baskets Heaters Steam rises Water in the base boils into steam

LORDERAN vertical lab autoclaves sterilize at up to 134 °C at a working pressure of 0.23 MPa (33 psi). Their non-vent or liquid mode keeps the vessel sealed so pressure falls slowly and liquids do not boil over.

Horizontal, gravity displacement

Steam enters from the top and pushes heavier cold air out through the drain.

Steam in Basket Cold air to drain Door No vacuum pump; steam is lighter than air

Prevacuum sterilizers use a vacuum pump instead, which removes air faster from porous and wrapped loads. In CDC's table, a prevacuum cycle for wrapped instruments needs 4 minutes at 132 °C against 15 minutes in gravity displacement.

SteamCold air pushed outHeating elements

Chamber sizes, drawn to scale

The listed volume is the chamber; the baskets are smaller. Size for the largest load you run in one cycle and leave space between items. University of Kentucky Research Safety advises against crowding items or overfilling bags and containers, and against sealing bottles tightly, so steam can reach every surface.

Vertical, top-loading

Chamber diameter × depth, mm

35 L300 × 45050 L350 × 51575 L400 × 585100 L450 × 620150 L500 × 747

Horizontal, front-loading

Chamber diameter × length, mm

150 L472 × 850200 L472 × 1130300 L620 × 960400 L620 × 1280

Both drawings use the same scale. Chamber sizes from the LORDERAN specifications; the 75 L size is offered in the drying and four-mode vertical models.

Vertical or horizontal

Vertical autoclave
Horizontal autoclave
Loading
Top-loading, stacked round baskets
Front door, long baskets slide in
LORDERAN sizes
35 to 150 L
150 to 400 L
Steam
Water in the chamber base, boiled by the heaters
Water supply at 0.1 to 0.2 MPa, heated inside the vessel
Power
Single-phase; 35 to 100 L in 110 V or 220 V, 150 L at 220 to 240 V
Three-phase 220 V or 380 V, 12 to 18 kW
Temperature
116 to 134 °C (106 to 134 °C on the four-mode model)
115 to 134 °C
Drying
Up to 240 min on the drying and four-mode models
Up to 999 min
LORDERAN price from
$1,999 (35 L)
$6,999 (150 L)

Cycles by load

Temperature and time depend on the load and on how the autoclave removes air. The figures below are minimum exposure times at temperature. Heat-up, exhaust and drying come on top, so a full cycle takes much longer than the hold.

Cycle
Notes
Wrapped instruments, gravity displacement
121 °C for 30 min, or 132 °C for 15 min
Then 15 to 30 min drying (CDC)
Wrapped instruments, prevacuum
132 °C for 4 min
Then 20 to 30 min drying (CDC)
Biohazardous waste
121 °C at 15 psi for at least 20 min
Total processing time 60 to 120 min; verify the autoclave monthly (University of Kentucky)
Liquids and media
Liquid cycle with a slow exhaust
Slow pressure release prevents boil-over; leave caps loose so steam can enter
Agar
Melt at 60 to 100 °C, hold at 45 to 60 °C
Mode M3 on the LORDERAN four-mode autoclave

Power and circuits

Check the circuit before you order. A vertical autoclave heats its water with 2.5 to 4.5 kW of elements, so the current at 120 V is high. Every LORDERAN vertical model offered in 110 V draws more than 20 A at 120 V.

At 220 V the same units draw 12 to 20.5 A. The 150 L vertical runs on 220 to 240 V only and draws 27 A. Horizontal autoclaves need a three-phase supply and a water connection.

Ask your electrician to confirm the circuit and plug type for the model you choose.

Current draw at 120 V, LORDERAN vertical autoclaves

Listed current for the 110 V versions. The dashed line marks 20 A.

35 L2.5 kW, bolt-lid and four-mode
21 A
50 L3 kW
25 A
75 and 100 L4.5 kW
37.5 A

The 35 L with drying uses 3 kW and draws 25 A at 120 V. Bars share one scale, starting at zero.

A three-question check

QUESTION 1Are you sterilizing sealed food packs such as retort pouches, cans or jars?
Yes
Counter-pressure food autoclaveCompressed air keeps packs from bursting. 100 or 150 L.
No
QUESTION 2Do you need more than 150 L per cycle, or a front-loading chamber for long baskets?
Yes
Horizontal autoclave, 150 to 400 LPlan for three-phase power and a water supply.
No
QUESTION 3Must loads come out dry, or must no steam be vented into the room?
Yes
Vertical with drying, or the four-mode modelThe four-mode model adds liquid and agar programs.
No
Bolt-lid vertical autoclave, 35 to 150 LSet temperature and time; it heats, holds and exhausts on its own.
VERTICAL, BOLT LID$1,99935 L, up to 150 L
VERTICAL WITH DRYING$2,39935 L, up to 100 L
HORIZONTAL$6,999150 L, up to 400 L

LORDERAN prices from, as of November 2026, with free delivery and duties paid to your loading dock.

LORDERAN autoclaves

35 to 400 LUp to 134 °C0.22 to 0.23 MPa working pressure110 V, 220 V or three-phase

Four vertical models and one horizontal model. The vertical autoclaves sterilize at up to 134 °C at a working pressure of 0.23 MPa (33 psi), with a sterilization timer of up to 999 minutes. The drying and four-mode models condense exhaust steam in a built-in tank or bottle, so no steam is vented into the lab.

Questions buyers ask

What size autoclave do I need?

Take the largest load you run in one cycle, put it in the baskets with space between items, and choose the chamber that holds it. If you run several loads a day, a larger chamber cuts the number of cycles. Compare basket sizes as well as the listed volume.

Can I autoclave liquids and media?

Yes, with a liquid cycle. The LORDERAN vertical autoclaves have a non-vent or liquid mode that lets pressure fall slowly so liquids do not boil over. Leave caps loose so steam can enter the bottles.

Do I need a 220 V circuit?

Not always, but every LORDERAN vertical model offered in 110 V draws more than 20 A at 120 V, from 21 A for a 35 L to 37.5 A for a 100 L. At 220 V they draw 12 to 20.5 A. Confirm the circuit with your electrician before you order.

What should never go in an autoclave?

University of Kentucky Research Safety lists volatile or corrosive chemicals, radioactive materials, sharps containers, and red-bag regulated medical waste.

How do I know a cycle worked?

CDC notes that the effectiveness of steam sterilization is monitored with a biological indicator. University of Kentucky requires autoclaves that decontaminate biohazardous waste to be verified monthly. Chemical indicator tape only shows that the item was exposed to heat.

Do I need a prevacuum autoclave?

Prevacuum cycles remove air faster from porous and wrapped loads, which shortens the hold, as in CDC's 4 minutes at 132 °C for wrapped instruments. Gravity displacement handles media, glassware, unwrapped items and waste, with longer holds.