How to Protect a Lab Autoclave from Voltage Problems

November 24, 2026

Last updated November 2026 · LORDERAN how-to guide

Short answer: a lab autoclave heats its water with 2.5 to 4.5 kW of elements, so the 120 V versions draw 21 to 37.5 A. On a long, shared or undersized circuit, the voltage at the outlet drops each time the heaters switch on. Low voltage is hard on the contactors and relays that switch those heaters: they close weakly, chatter and arc, and their coils run hot. Prevent it at the source with a dedicated circuit sized by an electrician, no extension cords or power strips, a voltage check at the outlet with the heaters running, and the 220 V version where the room allows it. Every LORDERAN sterilizer also has a built-in wide-voltage protection stage that buffers high and low swings before they reach the control power supply, relays and contactors.

What happens inside the autoclave

The controller runs on a small control power supply. When it calls for heat, it energizes the coil of a contactor or relay, and the contactor connects the heating elements to the supply. Every part in that chain depends on the voltage at the outlet, and that voltage is lowest at the moment the heaters draw current.

The wire takes its share

Current through the wire drops voltage before it reaches the plug.

panel100 ft of 10 AWGautoclave120 V120 VHeaters off: no drop114 VHeaters on, 25 A: 6 V lost in the wire

The contactor feels it first

A contactor's coil needs enough voltage to pull the contacts shut.

Rated coil voltageLow coil voltagestrong pull, sealedweak pull, bouncescontactsarcingCloses firmlyChatter and arcs
Heater current and contactsControl sideSupply wiring

What the outlet should read

In the US, ANSI C84.1 sets the voltage ranges. For a 120 V system, utilities design for 114 to 126 V at the meter (Range A). The standard then allows for the drop in the building's own wiring, down to 108 V at the equipment. Equipment is expected to work across that range.

The National Electrical Code recommends sizing branch circuits for no more than a 3 % voltage drop, and no more than 5 % for feeder and branch circuit together. Those are recommendations, not requirements, so older buildings and circuits added later do not always meet them.

IEEE 1159 names what happens outside the normal range. A sag is a drop below 90 % of nominal (108 V on a 120 V system) lasting from half a cycle to a minute. If it lasts longer than a minute it is a sustained undervoltage. A swell is a rise above 110 %.

A 120 V outlet, by the standards

Bands from ANSI C84.1 Range A and IEEE 1159.

104108114120126132Swell, above 132 VRange A at the meter114 to 126 VLost in building wiringdown to 108 V at the plugSag or undervoltagebelow 108 V (0.9 pu)volts, 120 V nominalan outlet as heaters switch on

Why low voltage wears relays and contactors

The contactor is a switch moved by an electromagnet. The magnet's pull falls as the coil voltage falls. IEC 60947-4-1 requires contactors to close reliably between 85 % and 110 % of their rated coil voltage, so a coil that sees less than that is outside what it was built for.

Schneider Electric and ABB, two of the largest contactor makers, describe the same chain of damage.

  1. 1
    A slow or incomplete closeA coil draws more current while its magnet is open than once it has closed. If low voltage keeps it from closing quickly, that high current lasts longer, the coil overheats and its insulation degrades. Schneider names undervoltage among the usual causes of repeated coil burnouts.
  2. 2
    Chatter and micro-openingsIf the voltage dips while the contactor is closed, the magnet may not hold the contacts firmly. They vibrate or open for an instant under full heater current.
  3. 3
    Arcing, pitting and weldingEach micro-opening draws an arc across the main contacts. The arcs pit the contact surfaces, and repeated arcing can weld them together. ABB notes that voltage fluctuations from long cable runs lead to chattering and welded contacts.
  4. 4
    High voltage hurts tooA coil held above its rating runs hot. Schneider notes a small rise is tolerated only if the temperature limit is not exceeded, and a long rise can cause failure. Brief spikes from lightning or switching, called surges, can also damage the control electronics.

Low voltage also slows the cycle. The power of a heating element falls with the square of the voltage. At 108 V a 120 V heater delivers about 81 % of its rated power, so heat-up and the full cycle take longer even when nothing fails.

How much the wiring loses

Voltage drop is current times the resistance of the wire, out and back. The 120 V versions of these autoclaves draw more current than a standard 15 or 20 A outlet can supply, which is why the circuit matters more for an autoclave than for most lab equipment.

Current draw, LORDERAN autoclaves

MODELHEATER POWERAT 120 V (LISTED)AT 220 V
35 L vertical2.5 kW21 Aabout 11 A
35 L vertical with drying3 kW25 Aabout 14 A
50 L vertical3 kW25 Aabout 14 A
75 and 100 L vertical4.5 kW37.5 Aabout 20.5 A
150 L vertical—not offered27 A (220 to 240 V)
Horizontal, 150 to 400 L12 to 18 kWnot offeredthree-phase 220/380 V

220 V figures for the 35 to 100 L models are calculated from rated power (current = power ÷ voltage). Check the rating plate of your unit.

Worked examples: voltage lost in the branch circuit

LOADCOPPER WIRE, ONE WAYVOLTS LOSTOF SUPPLY
50 L at 120 V, 25 A10 AWG, 50 ft3.1 V2.6 %
Same autoclave, same wire, 100 ft10 AWG, 100 ft6.2 V5.2 %
75 or 100 L at 120 V, 37.5 A8 AWG, 50 ft2.9 V2.4 %
50 L, 220 V version, about 14 A10 AWG, 100 ft3.4 V1.5 %

Single-phase drop = 2 × length × current × resistance. Resistance for stranded copper from NEC Chapter 9, Table 8: 1.24 Ω per 1,000 ft for 10 AWG and 0.778 Ω for 8 AWG. Wire size for your installation is your electrician's decision. The same autoclave on the same 100 ft run loses 5.2 % at 120 V and 1.5 % as a 220 V version, because half the current over the same wire drops half the volts out of twice the supply.

Prevention checklist

Most of these cost nothing once the autoclave is planned in. Electrical work is for a licensed electrician.

Before you order

  1. 1
    Ask what the room has120 V, 208 V or 240 V, and the breaker sizes free in the panel. Tell us before ordering so we can confirm the right version.
  2. 2
    Prefer the 220 V versionWhere the room can supply it, it draws about half the current of the 120 V version, so the wiring loses far less.
  3. 3
    Plan a dedicated circuitOne circuit for the autoclave alone, sized by your electrician for the listed current as a continuous load.

At installation

  1. 1
    Plug straight into the wallNo extension cords, power strips or adapters. Most are rated 15 A, below what these autoclaves draw, and every joint adds resistance.
  2. 2
    Measure under loadHave the electrician read the outlet voltage with the heaters on, not just at idle, and tighten every connection on the circuit. Keep the reading as a baseline.
  3. 3
    Add surge protection where it fitsUL 1449 Type 1 or Type 2 surge protective devices mount at the service entrance or panel. A plug-in surge strip is not rated for a 21 to 37.5 A autoclave.

In daily use

  1. 1
    Keep the circuit to itselfDo not add ovens, heaters or another autoclave to it later. Their switching causes sags too.
  2. 2
    Notice the signsBuzzing at heat-up, display resets or slower cycles all point to the supply. See the table below.
  3. 3
    If low readings persistLow at the meter: ask your utility to check the service voltage against ANSI C84.1. Frequent sags inside the building: an electrician can add an over and undervoltage monitoring relay.

What LORDERAN builds in

A relay that fails because of the supply looks, to the person standing at the autoclave, exactly like a faulty autoclave. The cycle stops, the load has to be run again and the lab waits for a part. Users should not have to diagnose their building's wiring to keep sterilizing.

So every LORDERAN sterilizer has a wide-voltage protection stage built in. It sits between the supply and the control side and buffers the hit when the supply runs too high or too low, so the control power supply, relays and contactors are not exposed to the full swing. It is standard on every LORDERAN sterilizer, not an option.

Wide-voltage here describes the protection, not the rating: run the autoclave at the voltage on its rating plate. The protection does not change the physics of the heaters, which still give less heat on a low supply, and it does not replace a correctly sized, dedicated circuit. Treat it as the last layer of protection, behind the site measures above.

Protection before the control side

Built into every LORDERAN sterilizer.

supply inWide-voltageprotectionControlpower supplyRelaysContactorsSwings are buffered before the control side

Warning signs

These usually point to the supply rather than the autoclave. Electrical checks are for a qualified electrician with the power off.

WHAT YOU NOTICELIKELY CAUSEWHAT TO DO
Buzzing or rattling from the cabinet when the heaters switch onA contactor coil getting too little voltage and chattering.Stop the cycle. Have an electrician measure the voltage at the outlet with the heaters on.
Display dims, resets or restarts mid-cycleA sag each time the heaters switch on, often from a shared or long circuit.Move the autoclave to a dedicated circuit and have the connections checked.
Heat-up takes longer than it used toLow voltage at the outlet. Heater power falls with the square of the voltage.Measure under load and compare with the readings from installation.
Breaker trips, or the plug or cord feels warmAn overloaded or undersized circuit, or a loose connection.Stop using the outlet and call an electrician.
The same relay or contactor fails again after replacementThe supply, not the part. A new part on the same supply wears out the same way.Measure the supply first, then contact LORDERAN with the model, serial number and the readings.

Manuals and guides

The user manual lists the electrical requirements for each model.

Buying guideLab autoclaves
How-to guideUsing an autoclave
ManualsSteam sterilizers
ProductsVertical autoclaves

Questions

Can I run an autoclave from an extension cord or power strip?

No. Most are rated 15 A, well below the 21 to 37.5 A these autoclaves draw at 120 V, and the extra length and joints add voltage drop and heat. Plug it straight into a dedicated outlet.

Will a surge protector fix low voltage?

No. A surge protective device clamps short spikes. It does nothing for sags or a supply that runs low. Those are fixed in the wiring: a shorter, heavier or dedicated circuit, or a utility check of the service voltage.

Does wide-voltage protection mean I can run a 220 V model on 120 V?

No. It protects the control power supply, relays and contactors against swings around the rated voltage. Always connect the autoclave to the voltage on its rating plate.

Should I choose the 120 V or the 220 V version?

If the room can supply 220 to 240 V on a suitable circuit, choose that version: it draws about half the current. Many US buildings have 208 V rather than 240 V, so ask your electrician which you have and tell us before ordering.

How do I find out whether my supply is the problem?

Have an electrician measure the outlet voltage with the heaters on. A reading well below 108 V on a 120 V circuit, or a large drop between idle and heating, points to the circuit. A power quality logger left on the circuit for a few days shows sags that a single reading can miss.