RPM vs RCF (× g): How to Convert and Choose a Centrifuge

November 21, 2026

Last updated November 2026 · LORDERAN buying guide

Short answer: protocols give centrifuge speed in × g because force, not rpm, does the separating, and the same rpm gives more force in a larger rotor. Convert with RCF = 1.118 × 10⁻⁵ × r × rpm², where r is the rotor radius in centimeters. To choose a centrifuge, find the highest × g your protocols call for and the largest volume per run, then pick the class that covers both.

Same rpm, different force

The force on a sample grows in proportion to its distance from the center of rotation. Run two rotors at the same speed and the sample in the larger one feels more force. That is why a protocol written as "4000 rpm" gives different results on different centrifuges, and "× g" does not.

Small rotor

Micro-tube rotor, sample about 62 mm from the center.

r≈ 1,100 × g4000 rpm, radius 62 mm

At 4000 rpm the sample feels about 1,100 times gravity. The same rotor reaches 30,642 × g at 21,000 rpm on the TGL-21M.

Large rotor

4 × 750 ml swing rotor, sample about 211 mm from the center.

r≈ 3,780 × g4000 rpm, radius 211 mm

At the same 4000 rpm the sample feels about 3,780 times gravity, more than three times as much, as on the DL-5B.

Radius rTubes and resulting force

The conversion

Relative centrifugal force (RCF) is the force on the sample as a multiple of Earth's gravity. It depends on the square of the speed, so doubling the rpm gives four times the force, while doubling the radius only doubles it.

Use the radius to the sample. Rotor data usually gives r max, the distance to the bottom of the tube, which is where pellets form.

RCF = 1.118 × 10⁻⁵ × r × rpm²r in centimeters, from the rotor axis to the sample
rpm = √( RCF ÷ (1.118 × 10⁻⁵ × r) )the same formula, solved for speed

Example: 8000 rpm at r = 12.7 cm gives 1.118 × 10⁻⁵ × 12.7 × 8000² ≈ 9,100 × g, the TDL-8A maximum.

RPM for common × g

Speed needed to reach each force at four rotor radii, rounded to the nearest 10 rpm. Check the result against your rotor's maximum speed before you run it.

TARGET
r = 6 cm
r = 10 cm
r = 15 cm
r = 20 cm
500 × g
2,730
2,110
1,730
1,500
1,000 × g
3,860
2,990
2,440
2,110
1,300 × g
4,400
3,410
2,780
2,410
2,000 × g
5,460
4,230
3,450
2,990
5,000 × g
8,630
6,690
5,460
4,730
10,000 × g
12,210
9,460
7,720
6,690
20,000 × g
17,270
13,380
10,920
9,460

Calculated with RCF = 1.118 × 10⁻⁵ × r × rpm². 6 cm is typical of micro-tube rotors; 15 to 20 cm of large swing rotors.

Worked example: serum tubes

BD states that its Vacutainer SST and PST gel tubes should be spun at room temperature at 1000 to 1300 RCF, for 10 minutes in a swing-bucket centrifuge or 15 minutes in a fixed-angle one. Here is that range on a TDL-5A swing rotor with r ≈ 15.7 cm.

LOWER LIMIT2,390 rpmGives 1000 × g at r = 15.7 cm.
UPPER LIMIT2,720 rpmGives 1300 × g at r = 15.7 cm.
TIME10 minSwing bucket. Allow 15 minutes in a fixed-angle rotor.

The TDL-5A and TGL-21M both have a Speed / RCF mode that converts for the selected rotor, so you can enter 1300 × g directly. Step by step: how to use a lab centrifuge.

How much force each class reaches

Low-speed benchtop and large-capacity centrifuges reach a few thousand × g: enough for blood tubes, cell pellets and bottles. High-speed models reach tens of thousands of × g in smaller tubes.

The force a centrifuge reaches depends on the rotor. Large buckets run slower than micro-tube rotors, so the top figure comes from the smallest rotor in each machine.

Maximum RCF by model

Blue: low speed and large capacity. Amber: high speed.

DL-5B / D5Mfloor, 4 × 750 ml
3,780 × g
TDL-5Abenchtop
4,390 × g
DL-5M / D5Bfloor, 4 × 1000 ml
4,640 × g
L580Rtabletop, 4 × 750 ml
5,060 × g
TDL-8Abenchtop
9,100 × g
TGL-21Mbenchtop high speed
30,642 × g
BH2200Rfloor high speed
52,000 × g
H2500R · GL-25M · BH2500Rfloor high speed
64,983 × g

One scale from 0 to 65,000 × g.

A two-question check

QUESTION 1Do any of your protocols call for more than about 10,000 × g?
Yes
High-speed centrifugeTGL-21M on the bench to 30,642 × g, or a floor model to 64,983 × g.
No
QUESTION 2Do you spin bottles of 750 ml or more, or many blood tubes per run?
Yes
Large-capacity centrifugeL580R on the bench, or DL-5B and DL-5M on the floor.
No
Low-speed benchtop centrifugeTDL-5A or TDL-8A, refrigerated, up to 9,100 × g.

LORDERAN centrifuges

Swing and angle rotorsLid lock−20 to +40 °CDatasheet per model

Every model locks its lid while running, and the refrigerated models hold −20 to +40 °C. The TGL-21M, L580R, DL series and floor high-speed models add an imbalance alarm. The DL-5B and DL-5M also come as ambient versions, D5M and D5B. See every model in the centrifuge collection.

Questions buyers ask

Should I set rpm or × g?

Set × g whenever the protocol gives it. The TDL-5A and TGL-21M convert between the two for the selected rotor; on any model you can work out the rpm with the formula or the table above.

Which radius do I use in the formula?

The distance from the rotor axis to the sample. Rotor tables usually give r max, to the bottom of the tube, where the pellet forms. A sample near the top of the tube feels less force.

Why do swing-bucket and fixed-angle times differ?

The tube sits at a different angle and the sample travels a different path. BD, for example, gives 10 minutes for its gel tubes in a swing-bucket centrifuge and 15 minutes in a fixed-angle one at the same force.

How fast can a large bottle rotor run?

Slower than a micro-tube rotor. On the TGL-21M the swing rotors take up to 4 × 750 ml at 4000 rpm, while its angle rotors run micro tubes to 21,000 rpm.

Do I need a refrigerated centrifuge?

Choose one when samples must stay cold or at a set temperature during the run. For samples that need no cooling, the D5M and D5B run at room temperature for less.